Local Operation
It means that all sequences have passed in the program tests. If the test fails, LCD will show
as follows.
PROG. XX Err.
REPORT 1, 2, 3,…….10
PROG. XX indicates the file number of the program failed, 1 to 10. In addition, 1, 2, 3...10
shown by LCD stand for the failed sequence numbers. During program chain testing, if the
failed program files are more than one set, you can use
▲
,
▼
to read the contents of failed
programs.
4.2.4 Setting the Specification
The
GO/NG
key is to enable/disable specification checking function. The Load will
compare measurement data with the set specifications for HIGH and LOW boundary when
the SPEC TEST is ON, and the LED, GO/NG, is lighted on the panel. To set specifications
for the load; please refer to the section 3.5 at the end of each mode for parameters setting.
The specification unit for CC, CR, and CP modes is volt while CV mode is current. There
are three levels for each mode: CENTER, HIGH and LOW. The CENTER level must be set
by the value of channel input reference level. The HIGH and LOW levels can be set by the
value or percentage selected in configuration
SPEC.ENTRY MODE.
The HIGH/LOW
percentage range is from 0 to 100%.
The default setting of HIGH and LOW is 100%. The CENTER value is half of the range.
Please refer to section 4.2.5 for the specification selection set by Value or Percentage.
4.2.5 Setting the Configuration
The Electronic Load provides useful features such as Von point, Current limit, Auto Load On,
etc. To use these powerful features, you must set relevant parameters in accordance with
application needs by the configuration setup. This procedure is only needed for initial setup
of a test operation. The configuration of the Load is stored in EEPROM. To set
configuration you must press the
CONF
key.
Set the voltage range for CC, CP mode
. There are two voltage ranges for CC or CP mode.
High range is for high voltage and low range is for low voltage so as to get better voltage
resolution. The default setting of V range is HIGH.
CC, CP V RANGE SELECT
1: HIGH 2: LOW
4-13
Содержание 63200 Series
Страница 1: ......
Страница 2: ......
Страница 10: ......
Страница 14: ......
Страница 27: ...General Information Model 63205 Model 63206 and 63207 1 13...
Страница 30: ......
Страница 38: ......
Страница 41: ...Operation Overview V MAX 80V Figure 3 2A The Rear Panel of 63201 63202 63203 63204 3 3...
Страница 78: ......
Страница 108: ...High Power DC Electronic Load 63200 Series Operation Programming Manual Return Example 2 5 7 26...
Страница 122: ......