Local Operation
4. Set Slew Rate
There are 250 discrete steps in each range. Set the rise 20 A/
µ
S and fall slew rates to
0.5A/
µ
S by pressing
2
,
0
,
ENTER
for rise and
0
,
.
,
5
ENTER
for fall slew rate.
CCH
: 20.0A/uS
CCH
: 0.5A/uS
5. Set Voltage spec
Set the high voltage spec. to 6 V by pressing
6
,
ENTER
.
CCH VOLTAGE SPEC.
HIGH : 6.000V
Set the low voltage spec. to 4 V by pressing
4
,
ENTER
.
CCH VOLTAGE SPEC.
LOW : 4.000V
Then the display will go to the first editing page again.
L
Notice
Spec. value can also be programmed by PERCENT under Configure setting. You have to
input the CENTER Voltage, and then enter the HIGH and LOW percent to the center voltage.
6. Set dynamic current mode.
Press
DYNA
key under CC mode, and select current range by
RANGE
key. Program
the high current level to 50A by pressing
5
,
0
,
ENTER
keys. Program the low current
level to 5A by pressing
5
,
ENTER
keys.
CCDL1: 50.000A
CCDL2: 5.000A
Then set the dynamic function period T1 to 0.1 mS, period T2 to 0.2 mS by pressing
0
,
.
,
1
,
ENTER
and
0
,
.
,
2
,
ENTER
. The range of dynamic period is from 0.025
mS to 30 Sec.
4-5
Содержание 63200 Series
Страница 1: ......
Страница 2: ......
Страница 10: ......
Страница 14: ......
Страница 27: ...General Information Model 63205 Model 63206 and 63207 1 13...
Страница 30: ......
Страница 38: ......
Страница 41: ...Operation Overview V MAX 80V Figure 3 2A The Rear Panel of 63201 63202 63203 63204 3 3...
Страница 78: ......
Страница 108: ...High Power DC Electronic Load 63200 Series Operation Programming Manual Return Example 2 5 7 26...
Страница 122: ......