Programmable Bidirectional DC Power Supply 62000D Series
Operating and Programming Manual
3-26
1.
V SET/ I SOURCE SET
has 2 selections which are NONE/Vref (0-10V).
When selecting Vref=10V
it means the DC power supply’s output
0~100% will map to 0~10V as Figure 3-41 shows.
Figure 3-41
2.
V MEAS/ I MEAS
has 2 selections which are NONE/Vref (0-10V).
When selecting Vref=10V
it means the DC power supply’s output
0~100% will map to 0~10V as Figure 3-42 shows.
Figure 3-42
3.
When using APG, in case the error is too large be sure to calibrate the
APG settings and measurements first.
4.
For APG pin assignments, please refer to
3.2.4.1.2 LAN
This DC power supply uses LAN to provide remote operation. The LAN address is required
for remote operation.
1.
The user needs to connect the network cable to the DC power
supply for auto detection.
2.
If the network cable is not connected properly, it may cause the DC
power supply screen to show abnormally. Turn off the DC power
supply to resolve the network cable problem and reboot it to clear
the abnormal screen.
Содержание 62000D Series
Страница 1: ......
Страница 2: ...Get more information by downloading Chroma ATE Solutions APP ...
Страница 8: ...vi ...
Страница 9: ...vii ...
Страница 10: ...viii ...
Страница 14: ......
Страница 18: ......
Страница 35: ...Overview 1 17 Figure 1 6 Dimension of 62000D 600 1200 1800 Models ...
Страница 36: ......
Страница 52: ......
Страница 54: ...Programmable Bidirectional DC Power Supply 62000D Series Operating and Programming Manual 3 2 Figure 3 2 ...
Страница 63: ...Manual Operation 3 11 Figure 3 13 Figure 3 14 ...
Страница 109: ...Manual Operation 3 57 _EXT ON SIGNAL EXT ON DISABLE EXT ON ENABLE Output Output Figure 3 85 ...
Страница 110: ......
Страница 117: ...Program Sequence 4 7 Figure 4 8 Figure 4 9 ...