FAQ
Question 1:
The displayed current varies between 0 mA and 3 mA?
Answer 1:
The measured current almost corresponds to the one of the no-load operation's
detection. As a result the displayed value varies between 0 mA (no-load
operation detection) and the minimum measurable current of 2 mA. Consider the
tolerance of +- 3 mA too.
Question 2:
My measuring instrument doesn't display any measuring values, although at least
a 3 mA current runs.
Answer 2:
A current exceeding 3 mA may run through instruments having only an inductive
or a capacitive load and no measured value would be displayed. Reason: in the
case of a low power factor, the active power lowers under the point of no-load
operation's detection and every measured value is reset.