Freeze-dryer Alpha 1-4 LDplus
Freeze-dryer Alpha 2-4 LDplus
5 Set-up and connection
36 / 102
Version 11/2006, Rev. 2.2 of 17/10/2016
• sb
Translation of the original operating manual
Pos : 103 /200 Chris t/361 GT-BA Labor-Pil ot ( PROJ EKTE)/ Alpha/Al pha 1- 4_2-4 LD plus/050 Aufst ellung und Ansc hl uss/ 050-0060 Vakuumpumpe Al pha 1-4_2-4 LD plus @ 35\ mod_1430380091629_68. doc x @ 258883 @ 2 @ 1
5. 4. 6
Vakuumpumpe
5.5
Vacuum pump
NOTE
It is absolutely essential to refer to the separate instruction manual of the
vacuum pump and exhaust filter (if applicable)!
The vacuum pump must be connected to the vacuum connection of the unit
and to the electrical socket at the back of the unit, which is marked
accordingly (see chapter 2.1.1 - "Functional and operating elements").
NOTE
The vacuum pump is supplied with power by the unit, but the maximum
current for the vacuum pump is limited. It is absolutely essential to refer to
the label of the electrical outlet for the vacuum pump (see the following
picture)!
If the current requirement of the vacuum pump is higher than the value that
is stated on the label, the pump must be supplied separately via an on-site
power socket.
1
Label indicating the
maximum current
Fig. 16: Indication of the maximum current for the vacuum pump (example)
The oil mist that escapes from the pump during operation must be retained
or discharged via an exhaust filter (oil mist separator).
• We strongly recommend using an oil mist separator. The filter prevents
the contamination of the air with oil mist.
• For discharging the oil mist, connect a suitable hose (RZ-2.5 and RC-6:
½", DUO 5 or DUO 10 ¾") to the exhaust connector of the vacuum
pump.
• The hose line must be laid in such a manner that any condensation
water cannot flow back into the pump. In the case of a rising hose line,
we recommend using a condensate trap (Woulff bottle or wash bottle).