8 Maintenance and service
Version 11/2006, Rev. 2.2 of 17/10/2016
• sb
63 / 102
Translation of the original operating manual
• Check the material regularly (at least once a month) for
– cracks
– visible damage of the surface
– pressure marks
– signs of corrosion
– other changes.
• Replace any damaged components immediately for your own safety.
• Immediately rinse off the accessories if any liquids that may cause
corrosion come into contact with them.
• Clean the accessories outside the freeze-dryer once a week or
preferably after each use.
Pos : 188 /010 Uni vers almodul e/ Leerz eile @ 0\ mod_1202116244500_0.doc x @ 114 @ @ 1
Pos : 189 /010 Uni vers almodul e/ Leerz eile @ 0\ mod_1202116244500_0.doc x @ 114 @ @ 1
Pos : 190 /200 Chris t/360 GT-BA Labor-Pil ot ( STAND ARDMO DULE)/080 Wartung und I nst andhaltung/ 080-0020 D esi nfektion von Tr oc knungs kammer und Z ubehör @ 25\ mod_1404984107851_68. doc x @ 184461 @ 2 @ 1
8.2
Disinfection of the drying chamber and accessories
• Use commercially-available disinfectants such as, for example, Incidur
,
Meliseptol
, Sagrotan
, Buraton
, or Terralin
(available at specialised
trade).
• The freeze-dryers and the accessories consist of various materials. A
possible incompatibility must be considered.
• Before using cleaning or decontamination agents that were not
recommended by us, contact the manufacturer to ensure that such a
procedure will not damage the freeze-dryer.
• Please contact us if you have any queries (see chapter 7.3 - "Service
DANGER
If dangerous materials (e.g. infectious and pathogenic substances) are
used, the freeze-dryer and accessories must be disinfected.