Freeze-dryer Epsilon 1-4 LSCplus
Freeze-dryer Epsilon 2-4 LSCplus
10 Technical data
132 / 148
Version 04/2014, Rev. 2.13 of 01/12/2021
• sb
Translation of the original operating manual
Physical data (without vacuum
pump)
Epsilon 1-4 LSCplus
Epsilon 2-4 LSCplus
EMC according to EN 55011:
Class B
Class B
Heat emission:
min. 620 W
max. 1,100 W
min. 1,100 W
max. 1,600 W
Regrigerant data
Epsilon 1-4 LSCplus
Epsilon 2-4 LSCplus
Refrigerant:
- Global warming potential (GWP):
- Filling quantity:
- Max. permissible pressure:
- CO
2
equivalent:
R1270
3
62 g
25 bar
< 0,01 t
R1270
3
64 g
25 bar
< 0,01 t
Refrigerant:
- Global warming potential (GWP):
- Filling quantity:
- Max. permissible pressure:
- CO
2
equivalent:
R170
3
6 g
25 bar
< 0,01 t
R170
3
60 g
25 bar
< 0,01 t
Equipment connections
Epsilon 1-4 LSCplus
Epsilon 2-4 LSCplus
Aeration:
Hose nozzle DN6
(outside diameter 10 mm max)
max. 0.2 bar overpressure
Hose nozzle DN6
(outside diameter 10 mm max)
max. 0.2 bar overpressure
Vacuum connection:
Small flange connection DN25KF
(ISO 28403, DIN 2861)
Small flange connection DN25KF
(ISO 28403, DIN 2861)
Mains input:
CEE 7/4 plug (type F)
CEE 7/4 plug (type F)
Data interface (LAN):
RJ 45
RJ 45
Special equipment: water cooling
system
Epsilon 1-4 LSCplus
Epsilon 2-4 LSCplus
Part number:
112993
112994
Water feed flow temperature:
+15°C to +20°C
+15°C to +20°C
Water return flow temperature:
25°C max.
25°C max.
Water temperature difference:
10K max. with +15°C feed flow
temperature
5K max. with +20°C feed flow
temperature
10K max. with +15°C feed flow
temperature
5K max. with +20°C feed flow
temperature
Cooling water pressure:
10 bar max.
10 bar max.
Differential pressure between feed
flow and return flow:
1.5 bar max.
1.5 bar max.
Volumetric flow rate:
0.13 m
3
/h max.
0.20 m
3
/h max.
Heat dissipated by cooling water:
1.1 kW
1.6 kW
Cooling water feed flow connection:
R3/4" with hose nozzle DN 13
R3/4" with hose nozzle DN 13
Cooling water return flow connection:
R3/4" with hose nozzle DN 13
R3/4" with hose nozzle DN 13
Option: cooling water flow controller
(factory setting):
12 bar (refrigeration unit, high
pressure side)
12 bar (refrigeration unit, high
pressure side)