Freeze-dryer Epsilon 1-4 LSCplus
Freeze-dryer Epsilon 2-4 LSCplus
7 Malfunctions and error correction
Version 04/2014, Rev. 2.13 of 01/12/2021
• sb
111 / 148
Translation of the original operating manual
Pos : 222 /200 Chris t/360 GT-BA Labor-Pil ot ( STAND ARDMO DULE)/070 St örungen und F ehl ersuche/ 070- 0010- 0020-0020 Bel üft ungsventil (z. B. E2-4) @ 25\mod_1404984751036_68. doc x @ 186099 @ 4 @ 1
→
7.1.2.2
Aeration valve
A malfunction of the aeration valve may have several causes. One
potential source are contaminants such as product residues within the
valve.
• Switch the freeze-dryer off and disconnect the mains plug.
• Clean the valve (see chapter 8.1.3 - "Aeration valve").
• Put the freeze-dryer into operation again.
If there is still a leakage, the freeze-dryer must be checked by qualified
specialist personnel (see chapter 7.3 - "Service contact").
Pos : 223 /010 Univ ers almodul e/ Leerz eile @ 0\mod_1202116244500_0.docx @ 114 @ @ 1
Pos : 224 /200 Chris t/360 GT-BA Labor-Pil ot ( STAND ARDMO DULE)/070 St örungen und F ehl ersuche/ 070- 0010- 0020-0025 Druckst euerv entil @ 25\m od_1404984063525_68. docx @ 184209 @ 4 @ 1
→
7.1.2.3
Pressure control valve
A malfunction of the pressure control valve may have several causes.
NOTE
The inspection of the valve must be carried out by qualified specialist
personnel (see chapter 7.3 - "Service contact").
Pos : 225 /010 Univ ers almodul e/ Leerz eile @ 0\mod_1202116244500_0.docx @ 114 @ @ 1
Pos : 226 /200 Chris t/360 GT-BA Labor-Pil ot ( STAND ARDMO DULE)/070 St örungen und F ehl ersuche/ 070- 0010- 0020-0030 G ummiv entile @ 25\mod_1404984064588_68. doc x @ 184223 @ 4 @ 1
→
7.1.2.4
Rubber valves
In order to identify a leaking rubber valve, the valves must be checked
individually:
• Remove the rubber valve and seal the connection at the drying
chamber with a rubber stopper.
• Check the tightness under vacuum until the leaking valve has been
localised.
• Clean the valve or replace it if necessary.
Pos : 227 /010 Univ ers almodul e/ Leerz eile @ 0\mod_1202116244500_0.docx @ 114 @ @ 1
Pos : 228 /200 Chris t/360 GT-BA Labor-Pil ot ( STAND ARDMO DULE)/070 St örungen und F ehl ersuche/ 070- 0010- 0020-0040 Vak uummess sonde @ 25\mod_1404984065691_68. docx @ 184237 @ 4 @ 1
→
7.1.2.5
Vacuum sensor
Vacuum sensors have a limited service life and can be ordered as spare
parts.
Capacitive vacuum sensors
Capacitive vacuum sensors may experience a measurement drift due to
long-term use, soiling or sudden aeration. In this case, the vacuum sensor
must be adjusted (see chapter 8.1.9 - " Vacuum sensor ").
Pos : 229 /010 Univ ers almodul e/ Seit enwechs el @ 0\mod_1202116244312_0.docx @ 105 @ @ 1