Freeze-dryer Epsilon 2-10D LSCplus
5 Set-up and connection
48 / 161
Version 07/2022, Rev. 3.0 of 20/07/2022
• sb
Translation of the original operating manual
5.2
Supply and disposal connections
→
5.2.1
Power supply
→
5.2.1.1
Connection
DANGER
Danger of life caused by electric shock
There is a risk of electric shock when touching current-carrying
components.
This may lead to ventricular fibrillation, cardiac arrest, or respiratory
paralysis.
• Only qualified electricians are authorised to perform work on the
electrical system of the freeze-dryer!
• The electrical equipment of the freeze-dryer must be checked at
regular intervals by a qualified electrician!
• Defects such as loose connections or burnt cables must be eliminated
immediately.
DANGER
The operating voltage on the name plate must correspond to the local
supply voltage
Christ freeze-dryers are units of safety class I. Epsilon 2-10D LSCplus units
have a five-wire power cord with a 5-pin CEE plug (see chapter 10 -
"Technical data").
CAUTION
Prior to connecting the unit to the power supply system, check the rotary
field at the electrical outlet with a suitable measuring instrument:
The rotary field must be clockwise!
→
5.2.1.2
Customer-provided fuses
The freeze-dryer must be protected by way of a customer-provided mains
fuse with the correct rating.