Freeze-dryer Epsilon 2-10D LSCplus
8 Maintenance and service
130 / 161
Version 07/2022, Rev. 3.0 of 20/07/2022
• sb
Translation of the original operating manual
System/
Component
Detail / Part
Inspection
Periodical Maintenance
Repair / Rebuild
Be
fo
re
s
ta
rt
Pe
r wo
rk
in
g
d
a
y
In
s
p
e
c
ti
o
n
Rep
la
c
e
m
e
n
t
O
p
e
ra
ti
o
n
Cal
ib
ra
ti
o
n
Stainless steel door
sight glasses: clean,
illumination ok
X
X
Acrylic glass door:
surface clean and
undamaged
X
X
Machinery
–
general
Switch cabinet
X
X
Leakages
X
X
Noise emission
above average
X
X
Dynamic gaskets:
diaphragmas
50 steril.
cycles
Static gaskets:
O-rings, clamp, KF
100 steril.
cycles
Wires: fastened
X
Cable trays: closed
X
Framework:
no corrosion
X
Remove corrosion
Aeration
system
Aeration filter
50 steril.
cycles
Vacuum
system
Pump: oil level
X
X
X
Pump: oil condition
(colour,
contamination)
X
X
X
oil change
Exhaust filter:
emission of oil mist
X
X
X
Exhaust filter: oil
level
X
X
Cleaning of the filter
Residues / particles
inside the housing
X
Oil-leakage under
the pump
X
X
after
10 years
Rotary shaft seal,
bearings, vanes,