9
Once the update is complete, the screen will light up and the Reader will
print out a notification on the successful update.
Connect the Reader to the computer with the Serial to USB cable and open
the Data Manager.
In the upper right corner click Change to write.
Click Clear Project to remove previously inscribed Test Items.
Click Open file and select the file Test Items.
Tick all the required Test items and click Write.
When the test items have successfully been written, unplug the Reader from
the computer, connect to power and restart the Reader. The Reader is now
ready to use.
Reader operation
Reading test strips
Turn on the Reader on the left-hand side of the Reader.
Select
Test
on the main screen.
Select
Qualitative
for antibiotic tests or
Quantitative
for aflatoxin tests.
OBS!! Quantitative tests require the standard curve corresponding to the
batch in use to be loaded before testing. See the Reader section step 10
under Getting Started for guidance.
Select the relevant test from the list of tests.
Fill out Sample ID, Lot No., and Operator ID (optional).
Press
Test
and the test port located to the right in the front of the Reader will
open and a “platform” will emerge.
As soon as the test incubation has finished as according to the test protocol,
place the test strip in the Reader cartridge.
Insert the test cartridge fully into the port on the Reader.
Press
Test
again, the cartridge will enter the Reader and the test is initiated.
The reading will start automatically and show the
Result
screen after a few
seconds.
Interpret the result: Positive results are shown with a
+
after the line
antibiotics type. Negative results are shown with a
-
after the line antibiotics
type. The numbers given represent the line ratio: the intensity of the test