6
Getting started
Reader
Unpack the Reader carefully and check that the following parts are present:
Reader
Power cable
Test cartridge
Touch screen pen
Thermal printer paper
Serial to USB cable
Converter
Place the Reader on a level and stable surface.
Open the printer cover, insert thermal printing paper, and close the printer
cover.
Connect the power supply.
Turn on the Reader (switch on the left side of the Reader).
After the system self-test the
Home
screen will appear.
Choose
Settings
and enter the initial password CH123456 to edit the
instrument elements. Press
OK
.
Select
Language
to change the language of the Reader. Press
Confirm
to
approve the change and
Back
to return to all setting options.
Select
Date & Time
and use the
▼
/
▲
buttons on the screen to adjust the
date and time fields to the current date and time. Press
Confirm
to accept
the changes or
Cancel
to return to the menu screen without setting the date
and time.
When reading quantitative tests, the standard curve from the specific batch
must be loaded. To load a curve go to
Settings
, select
Load a curve
, click
View
to see if any previous curves have been loaded. If yes, click
Empty
to
remove it. Insert the included USB from the test kit in use in the
Data In
slot
located in the front of the Reader and select
Load.
The curve is loaded. For
every new batch of a quantitative test, a standard curve must be loaded.
The MilkSafe® Desktop Reader is now ready for use.
1
2
3
4
5
6
7
8
9
10