line divided by the intensity of the control line. For qualitative tests, the ratio
numbers cannot be used for quantitative purposes.
Ratio < 0.9: Positive
Ratio 0.9 – 1.1: Weak positive
Ratio >1.1: Negative
Upon test completion, the cartridge will exit the Reader and can be removed.
The platform should be pushed back in, once the Reader is turned off.
The test result is automatically saved on the Reader.
The result can be printed with the built-in printer by selecting
.
To perform a new test press
Home
and repeat step 2 through 12.
Test records
To view previous test results, select
Records
.
Select
Qualitative
for antibiotic tests or
Quantitative
for toxin tests.
To scroll up and down, either use the
PgUp
and
PgDn
buttons or the scrolling
bar. To scroll to the side use the scrolling bar in the bottom.
To resize the columns drag on column header edges.
View the details of a specific test result by first selecting the test by clicking
anywhere on the line and then select
Details
.
Print the results by selecting
Print.
Verification procedure
Verify the accuracy of the measurements performed with the MilkSafe™ Desktop
Reader by reading standard strips. The standard strips with 4 lines can be used to
verify the results of MilkSafe™ 2BC while standard strips with 5 lines can be used
to verify the results of MilkSafe™ 3BTC, 3BTS and 4BTSC.
Select the standard strip corresponding to the test you are using.
Complete step 3 through 10 in section Reading test strips with the standard
strip. The strip is ready for use.
The obtained ratios for each substance line should be 1.0 ± 0.1. If the ratios
exceed 1.0 ± 0.2, perform the reading again. If the ratios again exceed
1.0 ± 0.2, please contact your local Chr. Hansen representative.