![Chison SonoEye P3 Скачать руководство пользователя страница 9](http://html.mh-extra.com/html/chison/sonoeye-p3/sonoeye-p3_user-manual_2602727009.webp)
Digital Color Doppler Palm Ultrasound System
4
NOTE
Image quality will effect by conducted disturbance. Please do not use the effected image. And put system in
a simpler electromagnetic compatibility environment to start working, if conducted disturbance occurred,
stop using it and change the position for acquiring a better image.
Mode of operation
Continuous Operation
For maximum safety, always follow these guidelines:
Do not remove the protective covers on the system. These covers protect users from hazardous
voltages. Cabinet panels must remain in place while the system is in use. A qualified electronic
technician must make all internal replacements.
Do not operate this system in the presence of flammable gases or anesthetics.
Notice upon Installation of Product
Separation distance and effect from fixed radio communications equipment: field strengths from fixed
transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur
radio, AM and FM radio broadcast, and TV broadcast transmitter cannot be predicted theoretically with
accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site
survey should be considered. If the measured field strength in the location in which the ultrasound system is
used exceeds the applicable RF compliance level as stated in the immunity declaration, the ultrasound
system should be observed to verify normal operation. If abnormal operation is observed, additional
measures may be necessary, such as re-orienting or relocating the ultrasound system or using an RF
shielded examination room may be necessary.
Use either power supply cords provided by or designated by CHISON. Products equipped with a power
source plug should be plugged into the fixed power socket which has the protective grounding
conductor. Never use any adaptor or converter to connect with a power source plug (e.g. three-prong-
to-two-prong converter).
Locate the equipment as far away as possible from other electronic equipment.
Be sure to use only the cables provided by or designated by CHISON. Connect these cables following
the installation procedures (e.g. wire power cables separately from signal cables).
WARNING:
Portable RF communications equipment (including peripherals such as antenna cables and external
antennas) should be used no closer than 30 cm (12 inches) to any part of this medical system, including
cables specified by the manufacturer. Otherwise, degradation of the performance of this equipment could
result.
Содержание SonoEye P3
Страница 30: ...Digital Color Doppler Palm Ultrasound System 25 ...
Страница 76: ...Digital Color Doppler Palm Ultrasound System 71 ...