Digital Color Doppler Palm Ultrasound System
57
between patients. It is very important, therefore, to minimize the risk of disease transmission by using
barriers and through proper processing between patients.
Risk of Infection
ALWAYS clean and disinfect the system between patients to the level appropriate for the type of
examination and use FDA-cleared system sheaths where appropriate.
Adequate cleaning and disinfection are necessary to prevent disease transmission. It is the responsibility of
the equipment user to verify and maintain the effectiveness of the infection control procedures in use.
System Cleaning process:
DO disconnect the system from the displayer prior to cleaning/disinfecting the system. Failure to do so could
damage the system.
Perform Cleaning system after each use
Disconnect the system and remove all coupling gel from the system by wiping with a soft cloth and
rinsing with flowing water.
Wash the system with mild soap in lukewarm water. Scrub the system as needed using a soft sponge,
gauze, or cloth to remove all visible residue from the system surface. Prolonged soaking or scrubbing
with a soft bristle brush (such as a toothbrush) may be necessary if material has dried onto the system
surface.
WARNING
To avoid electrical shock, always turn off the system and disconnect the system before cleaning the
system.
CAUTION
Take extra care when handling the lens face of the system. The lens face is especially
sensitive and can easily be damaged by rough handling. NEVER use excessive force when
cleaning the lens face.
Rinse the system with enough clean potable water to remove all visible soap residue.
Air dry or dry with a soft cloth.
CAUTION
To minimize the risk of infection from blood-borne pathogens, you must handle the system and
all disposables that have contacted blood, other potentially infectious materials, mucous
membranes, and non-intact skin in accordance with infection control procedures. You must