![Chison SonoEye P3 Скачать руководство пользователя страница 56](http://html.mh-extra.com/html/chison/sonoeye-p3/sonoeye-p3_user-manual_2602727056.webp)
Digital Color Doppler Palm Ultrasound System
51
Errors & Messages
Possible Cause
Solution
Power-indicating lamp is
not lit When user turn on
the system.
Type C cord may not be
connected, or may not be well
connected with the displayer.
Please contact with Sales
office, service department
and distributor.
Power indicating lamp is lit
When turn on the system,
but no images on displayer.
The restart time interval after
shutdown is too short
Wait 1 minute after power
off and then power on.
Menu bar displays on the
screen but no scanning
image.
Transmission frequency, gain
or STC control is not set
properly.
System is connected
improperly.
The system is in frozen status
Adjust the transmission
frequency, gain or STC
control.
Ensure the system is
connecting correctly
Defreeze the system by
pressing the FREEZE
icon.
Image quality is abnormal
Examination mode is not
correct.
The image post-processing
setting is abnormal.
Adjust image post-
processing settings or set
it to default.
The system is not working
properly
Internal circuit protection
Restart the system
9.5 Service Responsibility
If users install, use and maintain the system fully according to CHISON’s installation manual, operation
manual and service manual, then SonoEye P3 main unit has a life time of 5 years,
The warranty of the system after ex-work is as the time in the warranty card.
The system is a precise electronic system. Standard maintenance must be performed by CHISON’s
authorized service engineer during the life time of the product.