Press the corresponding button to transfer the exam mode between real time B-mode with Doppler mode
(with audio).
Sample along the whole length of the vessel. Ensure that the system is parallel to flow. Listen, then look,
when positioning the sample volume cursor.
Perform measurements and calculations, as necessary.
Record results with your recording devices.
Press
to resume imaging.
Repeat the above procedure until all relevant flow sites have been examined.
Replace the system in its respective holder.
5.1.8 B-Lines
Select the Lung preset to start an exam, press the lung icon on the left side while freeze image. The analysis
data of lung will display on the right bottom corner which include B-Lines, mean pleural thickness, maximum
pleural thickness, minimum pleural thickness, standard deviation on pleural and diffuse.
Edit pleural line:
Press the Edit icon on the left of image, press the image with single finger and blue line display, move the
blue line with finger to edit pleural line.
Add B-Lines:
Press Add icon on the left of image, a yellow line displayed, press it and move it to the target position and
release finger to add B-lines.
Delete B-Lines:
Press the existed B-Lines with single push to active current B-Lines, press the Delete icon on the left of
image area to delete current B-Lines.
Lung Report:
After finishing all operation on lung, press the Report key on side menu, select images on current exam and
compare exam to start lung analysis, and result will display below the image.
5.1.9 Biopsy Guide
Press Biopsy Guide icon and the guide line displays. Press the icon again to exit Biopsy Guide.
Adjust Biopsy Guide Line:
Press the Biopsy Guide line and slide it to adjust the guide line.
Adjust Biopsy Guide Angle:
Press the Biopsy Guide line, the angle icon displayed, press to slide it to adjust the Biopsy Guide angle.
5.1.10 SonoNeedle
Select SonoNeedle in B mode, the system will recognize and mark the needle position and needle trajectory