ARTIPRESS
28
13.4.1 Guidance and manufacturer’s declaration – electromagnetic immunity
FOR ALL ME EQUIPMENT
The ARTIPRESS-series ME EQUIPMENT is intended for use in the electromagnetic
environment specified below.
The customer or the user of the ARTIPRESS-series ME EQUIPMENT should assure that it
is used in such an environment.
Immunity test
IEC 60601
Test level
Compliance
level
Electromagnetic
environment –
guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
± 6kV contact
Complies
Floors should be wood,
concrete or ceramic tile.
If floors are covered with
synthetic material , the
relative humidity should be at
least 30 %.
± 8kV air
Complies
Electrical fast
transient/burst
IEC 61000-4-4
± 2kV per power
supply lines
Complies
Mains power quality should
be that of a typical
commercial or hospital
environment
± 1kV for
input / output
lines
Complies
Surge
IEC 61000-4-5
± 1kV line(s) to
line(s)
Complies
Mains power quality should
be that of a typical
commercial or hospital
environment
± 2kV line(s) to
earth
Complies
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
<5% UT (>95%
dip in UT ) for
0.5 cycles
Complies
Mains power quality should
be that of a typical
commercial or hospital
environment. If the user of
the ARTIPRESS-series ME
EQUIPMENT requires
continued operation during
power mains interruptions, it
is recommended that the
ARTIPRESS-series ME
EQUIPMENT be powered
from an uninterruptible power
supply or a battery.
40% UT (60%
dip in UT )
for 5 cycles
Complies
70% UT (30%
dip in UT )
for 25 cycles
Complies
<5% UT (>95%
dip in UT ) for 5
sec
Complies
Power frequency
(50/60 Hz) magnetic
field
IEC 61000-4-8
3 A / m
Complies
Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical
commercial or hospital
environment
NOTE : UT is the a.c. mains voltage prior to application of the test level.
13.4.2 Guidance and MANUFACTURER’S declaration – electromagnetic
IMMUNITY FOR ME EQUIPMENT THAT ARE NOT LIFE-SUPPORTING
The ARTIPRESS-series ME EQUIPMENT is intended for use in the electromagnetic
environment specified below. The customer or the user of the ARTIPRESS-series ME
EQUIPMENT should assure that it is used in such an environment.
Immunity test
Immunity test
Immunity
test
Immunity test
Portable and mobile RF communications equipment should be used no closer to any part
of the ARTIPRESS-series ME EQUIPMENT, including cables, than the recommended
separation distance calculated from the equation applicable to the frequency of the
transmitter.
Recommended separation
distance
Conducted RF
IEC 61000-4-6
3Vrms
150kHz to 80MHz
3V
( V
1
)
d =
P
V
1
5
,
3
Radiated RF
IEC 61000-4-3
3V/m
80MHz to 2,5GHz
3 V / m
( E
1
)
d =
P
E
1
5
,
3
80MHz to 800 MHz
d =
P
E
1
7
800MHz to 2,5 GHz
where P is the maximum output power rating of the transmitter in watts (W) according
to the transmitter manufacturer and d is the recommended separation distance in metres
(m).
Field strengths from fixed RF transmitters, as determined by an electromagnetic site
survey,
*1
should be less than the compliance level in each frequency range
.*2
Interference may occur in the vicinity of equipment marked with the following symbol:
NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies.