![Chiappa Arredamenti 120765 SLB RD Скачать руководство пользователя страница 47](http://html1.mh-extra.com/html/chiappa-arredamenti/120765-slb-rd/120765-slb-rd_instructions-for-use-manual_2597879047.webp)
Battery test
Check the condition of the backup battery by carrying out the following
test.
1. Disconnect the bed from the electricity supply.
2. Raise the mattress platform to maximum height
3. Raise the backrest and thigh sections as far as they will go.
4. Press and hold the CPR button and lower the mattress platform to
minimum height.
5. Apply maximum head down tilt (Trendelenburg).
6. Return the mattress platform to the level position.
7. Apply maximum foot down tilt (reverse Trendelenburg).
If this test is not completed successfully, connect the bed to the electricity
supply for at least eight hours to recharge the battery then perform the
test again. If the bed fails a second time, contact CHIAPPA ARREDAMENTI
S.r.l..
To maintain best performance see:
- chapter 7. Technical data / Electrical & Safety data / backup battery.
- chapter 7. Technical data / Transportation and storage / Caution.
IFU-120765-SLB-RD-Chiappa-Arredamenti-11/2017#ed04