SeaGauge-Remote G2
User’s Manual
version 031119
Brookings, OR 97415
45
www.chetcodigital.com
vDash via PC USB port
module only
FIRMWARE UPDATE SWITCH
SeaGauge-Remote firmware can be updated by loading new FASH IMAGES using a USB
connection to PC via the vDash application. Firmware file uploads can be performed only
when SW1 is in FIRMWARE UPDATE mode
Switch
Mode
SW1 = Left
Firmware update via USB
SW1 = Right
Port 0 or 1 in USB mode
When performing a firmware update via USB mode, first power off the unit and move the
switch UP (away from the case side and towards the center). Next connect the supplied USB
cable (8-pins on one end and TYPE A USB on other). Then re-power the unit and verify the
USB drivers are loaded and a virtual COM port is assigned to the unit. Once confirmed,
proceed to use vDash to connect and load firmware file.
Default position for calibration
programming using vDash
SeaGauge Firmware update position only
Once completed remove power and return SW1 to normal USB mode RIGHT (towards the
the center).