Chemtrac UV254 Monitor
– UVM5000 v 2.1
Chemtrac, Inc.
33
14.4
Ribbon Cable R eplacement
The blue-ribbon cable that connects the main circuit
board to the sensor board located on the rotating
assembly is a Teflon, high flex 10 conductor cable. It is
designed to withstand long term exposure to the deep UV
light present inside the cabinet.
However, even this high flex cable will need to be
replaced from time to time. Chemtrac recommends
replacing the blue-ribbon cable every 12-18 months as
preemptive maintenance.
The replacement is a very simple procedure requiring no
tools:
1. Unclip the end connectors that secure the old
ribbon cable to the control board and rotating arm
by opening the wing clips.
2. Pull out the connector and reconnect the end
connectors of the ribbon cable.
3. Close the wing clips to secure the new cable.
Note:
A spare cable is included with every new UVM5000
Organics Monitor and is attached to the inside of the
cabinet door. To order more cables please contact
Chemtrac.