Chemtrac UV254 Monitor
– UVM5000 v 2.1
Chemtrac, Inc.
18
9
Menu Structure
MENU ITEM
DESCRIPTION
AVERAGING
On: 1 Off: 0
Average
Press
‘Enter’ then up/down to select running average for last 10 readings (Default = 0: Off)
UVT/UVA MODE
0 : UVT
1 : UVA
2 : Crltn.
Select
Press ‘Enter’ then up/down to select UVT, UVA, or Correlation mode (Default = 0: UVT)
SETPOINTS
UVT Warning
Press ‘Enter’ then up/down to select UVT warning set point (Default = 75)
UVT Alarm
Press ‘Enter’ then up/down to select UVT alarm set point (Default = 50)
UVA Warning
Press ‘Enter’ then up/down to select UVA warning set point (Default = 0.20)
UVA Alarm
Press ‘Enter’ then up/down to select UVA alarm set point (Default = 0.30)
Crltn. Warning
Press ‘Enter’ then up/down to select the correlation warning set point (Default = 160)
Crltn. Alarm
Press ‘Enter’ then up/down to select the correlation alarm set point (Default = 200)
4-20mA SETUP
4mA UVT
Press ‘Enter’ then up/down to select 4mA UVT reading for scaling (Default = 0)
20mA UVT
Press ‘Enter’ then up/down to select 20mA UVT reading for scaling (Default = 100)
4mA UVA
Press
‘Enter’ then up/down to select 4mA UVA reading for scaling (Default = 0)
20mA UVA
Press ‘Enter’ then up/down to select 20mA UVA reading for scaling (Default = 1)
4mA Crltn.
Press ‘Enter’ then up/down to select 4mA correlation reading for scaling (Default = 0)
20mA Crltn.
Press ‘Enter’ then up/down to select 20mA correlation reading for scaling (Default = 200)
SECONDARY DATA
Lamp 1
Displays the lamp output after it passes through the flow cell 1 cm position
Water 1
Displays the water output after it passes through the flow cell 1 cm position
VIS 1
Displays the raw VIS as it passes through the flow cell 1 cm position
Lamp 2
Displays the lamp output after it passes through the flow cell 2 cm position
Water 2
Displays the water output after it passes through the flow cell 2 cm position
VIS 2
Displays the raw VIS as it passes through the flow cell 2 cm position
Temperature
Displays temperature inside cabinet
– Celsius
Humidity
Displays the relative humidity inside cabinet
– RH
Status Code
Displays status code for system diagnostics
DRY CONTACT SETUP
See Set Alarms
0: Nothing
1: Alarm
Select
Press ‘Enter’ then up/down to select 0 or 1 to turn on/off dry contact (Default = 0: Nothing)