Chemtrac UV254 Monitor
– UVM5000 v 2.1
Chemtrac, Inc.
17
8.2
Using Known Water Method
Carry out these steps:
1. Press the Zero button on the monitor. The Calibration menu appears.
2.
Scroll down to ‘Custom Zero’ using the down arrow button. Press Enter.
3. A default v
alue is displayed in the ‘New UVT:’ field (in this example, 75.0).
4. Press the up arrow to increase the number above 75.0 or the down arrow to decrease the number
below 75.0. When you have keyed in the number of known water transparency, press Enter.
Note that if you only know your UVA value then this can be converted to UVT using the formula:
UVT = 100 * 10
(-UVA)
5. The monitor is now calibrated to the correct value.