16

Equipment Needed to Upgrade your V-3000:

!

PC with Windows XP, Windows Vista, Windows 7, or Windows 10.

NOTE

: Administrator rights will be required.

!

PC Serial interface cable (6 pin to 9 pin, supplied with instrument) and if necessary, USB Cable
Adapter (sold separately, Cat. No. A-0307).
OR
Power LabStation (sold separately, Cat. No. A-0302) with its serial interface cable and if necessary,
USB Cable Adapter (Cat No. A-0307).

Procedure for Firmware Revision Update:

Important Message!

  Before beginning the update:

!

Ensure that the V-3000 has a stable uninterrupted power supply (i.e. fresh batteries or Power
LabStation with power cord).

!

Close all other applications on your PC.

NOTE

: Updates do not delete user generated custom methods.

1.

Click on the link provided on the CHEMetrics website to open the update program.

2. Files have been compressed into a zip archive (*.zip).  Admin rights will be required to unzip the

archive and install the program on your PC.  Unzip the files and save them to your PC.  Open the
folder on your PC called “V3000_update”, then double click on the executable (*.exe) file.  When the
installation has occurred successfully, the executable icon (shortcut to the V-3000 upload program)
will be saved to your desktop.  

NOTE:

The step-by-step procedure for installing the program and unzipping the files varies among different
operating systems and individual PCs.  Consult with an IT Specialist for detailed instructions. 

3. With the V-3000 power off, connect the V-3000 to the computer using the supplied PC interface

serial cable.  If only USB ports are available, connect the V-3000 using the supplied serial cable and
USB Cable Adapter (sold separately, Cat. No. A-0307).

4. Turn the V-3000 on by pressing the ON/OFF key.

NOTE:

It does not matter which screen the instrument displays at the start of this update.

5. Double click on the V-3000 executable icon on your desktop.  Click “OK” to initiate the V-3000

Firmware Update.  The entire process will take approximately 10 minutes.  The V-3000 screen will
go through a series of programming tasks.

NOTE:  Do NOT click Cancel, remove PC cable, interrupt power supply, or attempt to stop the update.

This can result in meter malfunction requiring repair by the manufacturer.

6. When the update is complete, the PC screen will display “Firmware update successfully completed!”

as well as meter and revision information including the new Software and Program Version numbers. 
If the update fails, follow the prompts on the PC screen to re-initiate the upload.  When prompted,
carefully enter the V-3000 serial number.

NOTE:

If the update repeatedly fails, disconnect the V-3000 from the COM port and contact
[email protected].

7. When the update is completed successfully, disconnect the V-3000 from the COM port. 
8. Click File\Exit to close the program.  Do NOT click “OK”.

Содержание V-3000

Страница 1: ...OPERATOR S MANUAL V 3000 PHOTOMETER CHEMetrics Inc 4295 Catlett Road Midland Virginia 22728 Toll Free 1 800 356 3072 Phone 540 788 9026 Fax 540 788 4856 Web www chemetrics com Jan 18 Rev 8...

Страница 2: ...3 1 Turning the Photometer On 8 3 2 Inserting a Cell 8 3 3 Operating Modes 9 3 4 Navigation 9 3 5 Menu Overview 10 4 Photometry 4 1 Measuring the Concentration 11 4 2 Analysis Timer 12 4 3 Timer 12 4...

Страница 3: ...ic Range 0 200 to 2 000 absorbance units Operating Temperature 0EC to 50EC Humidity 75 Yearly Mean up to 30 days at a maximum of 95 Waterproof IP67 Power Supply 4 X 1 5 V AA approximately 5000 measure...

Страница 4: ...entration direct read mode toggle through available chemical forms for a particular analyte example NH3 NH3 N NH4 UNIT key In Photometry Concentration direct read mode toggle through available reporti...

Страница 5: ...g l Measured value with unit 15 K1503 NH3 N Program no Catalog no and Chemical form 13 mm 0 00 7 00 mg l Diameter of the cell and measuring range 11 30 13 15 12 Status line with date and time 1 5 Sock...

Страница 6: ...Y 436 0 40 0 Ferric Thiocyanate Chlorine 32 K 2513 13 N 517 0 5 00 DPD Chlorine 33 K 2523 13 N 517 0 5 00 DPD Chlorine Dioxide 37 K 2703 13 N 517 0 11 00 DPD Chromate 42 K 2803 13 N 557 0 3 50 as CrO...

Страница 7: ...0 as NO3 Cadmium Reduction Nitrite 125 K 7003 13 N 517 0 1 00 as N Azo Dye Formation Ozone 133 K 7423 13 N 517 0 5 00 DPD Oxygen 140 K 7503 13 N 610 0 2 00 Indigo Carmine Oxygen 141 K 7513 13 N 517 0...

Страница 8: ...hargeable Battery Pack PC Interface Cable 9 pin to 9 pin The Power LabStation provides the following features When the photometer is placed in the Power LabStation which is plugged into an electrical...

Страница 9: ...when the photometer is not in use To install batteries carefully loosen the two screws on the battery compartment on the back of the instrument Remove the battery compartment cover and insert the four...

Страница 10: ...the dust cover is pushed upward Insert a 28 mm cell into the sample compartment using downward pressure to ensure that it is fully seated Inserting a Vacu vials Ampoule 13 mm Requires installation of...

Страница 11: ...ed in the status line of the display There are choices for date format presentation day month year dd mm yy or month day year mm dd yy When storing measured values and calibrating the current date and...

Страница 12: ...ay RS232 Download Data Filter Filter ID Program Date Delete of 100 occupied i Filter No filter i Display Illumination Auto off On Off Contrast 0 100 Brightness 0 100 Reset Restore basic settings Inter...

Страница 13: ...d to i perform the zero adjustment The prompts are highlighted across the bottom of the display To proceed to each next prompt press START ENTER NOTE To perform the zero adjustment for Vacu vials Test...

Страница 14: ...s line of the display When the selected time expires the meter beeps Press START ENTER to stop the beeping 4 4 Measuring Diluted Samples If the concentration of a sample exceeds the measuring range of...

Страница 15: ...ured value memory capacity is full the meter will offer the option to overwrite the oldest stored measurement with the new storing command Another option is that all stored values can be deleted using...

Страница 16: ...1 55 06 0 15 NH3 N 4 50 mg l VALID Description of data separated by semi colons in CSV format Data storage number Date of storing Time of storing User assigned ID Program number only for the concentra...

Страница 17: ...fer Click on Capture Text Enter the path and name of the file created in Step 1 Click on Start 12 Make sure that the desired V 3000 output format has been chosen See Section 6 1 13 To download individ...

Страница 18: ...es varies among different operating systems and individual PCs Consult with an IT Specialist for detailed instructions 3 With the V 3000 power off connect the V 3000 to the computer using the supplied...

Страница 19: ...Error uploading method update Never press cancel during a method update Contact CHEMetrics for assistance Program number invalid Program number not recognized Only use program appropriate program numb...

Отзывы: