6.3 Downloading Data to a Computer

V-3000 Data Management Software for Data Exchange (Catalog Number A-0301) is available for easy data
downloading and manipulation. However, without the additional software, data can also be sent to a terminal
emulator program and then exported to a spreadsheet or database software application.

To download data without the V-3000 PC Software: 

NOTE

: These instructions are specific to the terminal emulator program HyperTerminal. 

1. Create and save a text file (*.txt) to receive the V-3000 logged data. Close this file.
2. With the V-3000 power off, connect the V-3000 to a computer using the supplied PC interface cable.
3. Turn the meter on by pressing the 

ON/OFF

 key.

4. Press the 

MENU

 key. The Configuration menu is opened.

5. Use the 

 or 

 keys to highlight System. Press 

START/ENTER

. The System menu is opened.

6. Use the 

 or 

 keys to highlight Measured value memory. Press 

START/ENTER

.

7. Open HyperTerminal by going to the Start button on your computer, then clicking on Programs/

Accessories/Communications/HyperTerminal.

8. A “New Connection” window will open. Enter a name for the terminal program, for example “V3000".

Select an icon to represent the terminal program.

9. A “Connect To” window will open. Select the Com (serial) port to which V-3000 is connected.

10.  Enter the port transmission settings:

Baud Rate:

1200, 2400, 4800, or 9600 

Data Bits:

 8

Parity:

None

Stop Bit:

1

Start Bit:

1

Flow Control:

Hardware

NOTE

: The baud rate selected in Hyper Terminal must match the V-3000 baud rate. V-3000 baud rate is set in

the Configuration/System/Interface/Baud rate menu. 

11. From the HyperTerminal menu, select Transfer. Click on Capture Text. Enter the path and name of the

file created in Step 1. Click on Start.

12. Make sure that the desired V-3000 output format has been chosen (See Section 6.1).
13.

To download individual readings

:  Use the 

 or 

 keys to highlight Display. Press 

START/ENTER

.

Readings will display one at a time. Use 

 or 

 keys to scroll. Press 

PRT

 key to download individual

readings in selected output format. 

To download entire datasets

: Use the 

 or 

 keys to highlight RS232 download. Press

START/ENTER

. This will download entire data sets in the selected output format according to the filter

criteria that has been set (See Section 5.3).

14. Downloaded readings should appear in the Hyper Terminal window. Once transmission is complete,

click on Transfer/Capture Text and select Stop. Exit the HyperTerminal program and disconnect the V-
3000 from the computer. 

15. The text file, which captured the V-3000 data, can now be opened in the appropriate software

application. Data may be reformatted as needed for spreadsheet or database manipulation.

6.4 Uploading Program Revisions

The V-3000 program version can be updated by visiting the CHEMetrics website, www.chemetrics.com and
clicking on V-3000 Method Update under the Support tab. The CHEMetrics website will inform customers
when new programs are available or existing programs have been updated. Also, product specific test kit
labels will alert the user about program updates as they occur. To ensure that the most current calibrations
are being used, it is recommended that every V-3000 user check the CHEMetrics website frequently.

To determine what firmware version is currently on your V-3000, turn the photometer on.  Press the Menu
key and select “Info”.  The info screen displays Software version and Program version (as well as Serial
number and main PC board identification).

Содержание V-3000

Страница 1: ...OPERATOR S MANUAL V 3000 PHOTOMETER CHEMetrics Inc 4295 Catlett Road Midland Virginia 22728 Toll Free 1 800 356 3072 Phone 540 788 9026 Fax 540 788 4856 Web www chemetrics com Jan 18 Rev 8...

Страница 2: ...3 1 Turning the Photometer On 8 3 2 Inserting a Cell 8 3 3 Operating Modes 9 3 4 Navigation 9 3 5 Menu Overview 10 4 Photometry 4 1 Measuring the Concentration 11 4 2 Analysis Timer 12 4 3 Timer 12 4...

Страница 3: ...ic Range 0 200 to 2 000 absorbance units Operating Temperature 0EC to 50EC Humidity 75 Yearly Mean up to 30 days at a maximum of 95 Waterproof IP67 Power Supply 4 X 1 5 V AA approximately 5000 measure...

Страница 4: ...entration direct read mode toggle through available chemical forms for a particular analyte example NH3 NH3 N NH4 UNIT key In Photometry Concentration direct read mode toggle through available reporti...

Страница 5: ...g l Measured value with unit 15 K1503 NH3 N Program no Catalog no and Chemical form 13 mm 0 00 7 00 mg l Diameter of the cell and measuring range 11 30 13 15 12 Status line with date and time 1 5 Sock...

Страница 6: ...Y 436 0 40 0 Ferric Thiocyanate Chlorine 32 K 2513 13 N 517 0 5 00 DPD Chlorine 33 K 2523 13 N 517 0 5 00 DPD Chlorine Dioxide 37 K 2703 13 N 517 0 11 00 DPD Chromate 42 K 2803 13 N 557 0 3 50 as CrO...

Страница 7: ...0 as NO3 Cadmium Reduction Nitrite 125 K 7003 13 N 517 0 1 00 as N Azo Dye Formation Ozone 133 K 7423 13 N 517 0 5 00 DPD Oxygen 140 K 7503 13 N 610 0 2 00 Indigo Carmine Oxygen 141 K 7513 13 N 517 0...

Страница 8: ...hargeable Battery Pack PC Interface Cable 9 pin to 9 pin The Power LabStation provides the following features When the photometer is placed in the Power LabStation which is plugged into an electrical...

Страница 9: ...when the photometer is not in use To install batteries carefully loosen the two screws on the battery compartment on the back of the instrument Remove the battery compartment cover and insert the four...

Страница 10: ...the dust cover is pushed upward Insert a 28 mm cell into the sample compartment using downward pressure to ensure that it is fully seated Inserting a Vacu vials Ampoule 13 mm Requires installation of...

Страница 11: ...ed in the status line of the display There are choices for date format presentation day month year dd mm yy or month day year mm dd yy When storing measured values and calibrating the current date and...

Страница 12: ...ay RS232 Download Data Filter Filter ID Program Date Delete of 100 occupied i Filter No filter i Display Illumination Auto off On Off Contrast 0 100 Brightness 0 100 Reset Restore basic settings Inter...

Страница 13: ...d to i perform the zero adjustment The prompts are highlighted across the bottom of the display To proceed to each next prompt press START ENTER NOTE To perform the zero adjustment for Vacu vials Test...

Страница 14: ...s line of the display When the selected time expires the meter beeps Press START ENTER to stop the beeping 4 4 Measuring Diluted Samples If the concentration of a sample exceeds the measuring range of...

Страница 15: ...ured value memory capacity is full the meter will offer the option to overwrite the oldest stored measurement with the new storing command Another option is that all stored values can be deleted using...

Страница 16: ...1 55 06 0 15 NH3 N 4 50 mg l VALID Description of data separated by semi colons in CSV format Data storage number Date of storing Time of storing User assigned ID Program number only for the concentra...

Страница 17: ...fer Click on Capture Text Enter the path and name of the file created in Step 1 Click on Start 12 Make sure that the desired V 3000 output format has been chosen See Section 6 1 13 To download individ...

Страница 18: ...es varies among different operating systems and individual PCs Consult with an IT Specialist for detailed instructions 3 With the V 3000 power off connect the V 3000 to the computer using the supplied...

Страница 19: ...Error uploading method update Never press cancel during a method update Contact CHEMetrics for assistance Program number invalid Program number not recognized Only use program appropriate program numb...

Отзывы: