Page 43
5. Overwrite the current Ic-value with the new
calculated Ic-value for this index (figure
). To do
so, press and hold
to quickly scroll to the
required number and either press it for 3 seconds
or press
to confirm the selected number (figure
Moving back:
Press
to switch to another input level.
To move back, press
.
6. Navigate to the next index by pressing
twice.
7. Repeat steps 5 and 6 for the rest of the calculated
indices, leaving index7 (Idx:[ 7]) untouched.
8. Press
to switch to another input level and
to leave the
Humidity
menu.
»
The recalibration has been saved.
9. Press
to leave the
Calibrate
menu.
10. Press
to leave the
Options
menu.
11. Press
to leave the main menu.
12. In case of an incorrect recalibration, it is possible to reset the device to its factory
"10.11 Resetting the device to its factory settings"
).
»
By restoring the factory calibration, recalibrations that have already been carried
out successfully are deleted.
76
77