Checkline humimeter RH5 Скачать руководство пользователя страница 41

Page 41

probe.

 

»

If the deviation revealed is more than 1.5 % relative humidity, it is recommended 
to perform a recalibration.

12.3  Performing a recalibration

To do so:

 The deviation revealed is more than 1.5% relative air humidity. All of the 

options must be activated (see 

"10.4 Activating options"

).

12.3.1  Determination of calibration values

To do so:

 Microsoft Excel calculation sheet required (available via email, on request by 

phone or by email to [email protected]).

1.  Press 

 twice or hold for 2 seconds.

2.  Select 

Options

. To do so, press 

 or 

 and confirm by pressing 

.

3.  Select 

Calibrate

. To do so, press 

 or 

 and confirm by pressing 

.

4.  Select 

Humidity

. To do so, press 

 or 

 and confirm by pressing 

.

 

»

The display will then show the first index, Index1 
(Idx:[ 1]) (figur

71

).

5.  Enter the indices 1 to 6 of the Ic values already sto-

red in the device into the corresponding fields of 
the MS Excel calculation sheet, paying attention to 
the volt values (figure 

72

). Navigate through the in-

dices by pressing 

 twice until you have returned 

to Index1 (Idx:[ 1]).

6.  Press 

.

7.  The Ic value (upper line) has to change (figure 

73

).

 

»

If Index7 (Idx:[ 7]) is displayed after the key is 
released, the key has been pressed too long.

 71 

 72 

 73 

Содержание humimeter RH5

Страница 1: ...bs 100 4 g m2 09 m s 4 90 Ugl 1 Operating Manual Always read this operating manual before using the humimeter RH5 RH5 1 humimeter RH5 RH5 1 Paper moisture meter with sword sensor for measuring the relative air humidity of paper Moisture meter Version 3 1 ...

Страница 2: ... Operating Manual Page 2 Your humimeter RH5 RH5 1 at a glance The main unit No Name 1 Air humidity and temperature sensor 2 Sword sensor 3 USB Port optional for RH5 4 Rubber protection cover 5 Keypad 6 Display 2 3 4 1 5 6 ...

Страница 3: ...main unit No Name 1 Sword sensor 2 Battery compartment The display No Name 1 Product type 2 Relativ air humidity in see 7 1 Definition of relative humidity and dew point 3 Display symbols 4 Temperature display 1 2 4 3 2 1 ...

Страница 4: ... data Enter letters View measurements Continue go right Delete measurements Left On off button display light Yes Save measured value Auto save Hold function The menus The device has three different menus product selection Data Log and main menu Product selection menu No Name 1 Change menu 2 Display illumination device on off 3 For changing the product type 2 1 3 ...

Страница 5: ...s Main menu The main menu comprises the following menu items Edit Logs Manual Logs Auto Logs Clear Logs Print Logs Last Log All Logs Clear Logs Send Logs Manual Logs Auto Logs Clear Logs Options Bluetooth Date Time Log Time Language Unlock C F BL On Time Auto Off Time Calibrate Materialcalib Online Send Password Reset Status 2 1 3 4 ...

Страница 6: ...0 1 4 Customer service 10 2 For your safety 11 2 1 Proper use 11 2 2 Improper use 11 2 3 User qualifications 11 2 4 General safety information 12 2 5 Warranty 12 3 On receipt of your device 12 3 1 Taking the device out of its packaging 12 3 2 Making sure that all of the components have been included 12 3 2 1 RH5 s scope of supply 13 3 2 2 RH5 1 s scope of supply 13 3 3 Inserting batteries 14 4 Usi...

Страница 7: ...save function time based 23 6 3 1 Activating the auto save function in the Options menu 23 6 3 2 Auto save function Saving measured values 23 6 4 Viewing individual readings 24 6 5 Viewing individual readings from a series of measurements 25 6 6 Deleting all measured values data log 25 6 7 Deleting individual measurement series 26 6 8 Deleting individual values from a single series of measurements...

Страница 8: ...d 36 10 11 Resetting the device to its factory settings 37 11 Cleaning and maintenance 38 11 1 Changing the batteries 38 11 2 Care instructions 38 11 3 Cleaning the device 38 12 2 point optionally 3 point calibration 39 12 1 Assembling the calibration equipment 39 12 2 Determining the deviation 40 12 3 Performing a recalibration 41 12 3 1 Determination of calibration values 41 12 3 2 Entering the ...

Страница 9: ... device 1 2 Limitation of liability All of the information and instructions provided in this operating manual have been compiled on the basis of the current standards and regulations the state of the art and the extensive expertise and experience of Schaller GmbH Schaller GmbH does not accept any liability for damage associated with the following which also voids the warranty Non observance of thi...

Страница 10: ...damage to property or equipment i Information This symbol indicates important information that enables users to use the de vice more efficiently and cost effectively 1 4 Customer service For technical advice please contact our customer service department at Schaller GmbH Max Schaller Straße 99 A 8181 St Ruprecht an der Raab Telephone 43 0 3178 28899 Fax 43 0 3178 28899 901 E Mail info humimeter co...

Страница 11: ...ty of stacks of cardboard and paper and for determining the residual humidity of diverse materials Easy to use device for automatic climate monitoring of printing rooms or storage rooms 2 2 Improper use The device must not be used in ATEX The device is not waterproof and must be protected from water and fine dust 2 3 User qualifications The device must only be operated by people who can be expecte...

Страница 12: ... Warranty The warranty does not apply to Damage resulting from non observance of the operating manual Damage resulting from third party interventions Products that have been used improperly or modified without authorisation Products with missing or damaged warranty seals Damage resulting from force majeure natural disasters etc Damage from improper cleaning Batteries older than six months 3 On rec...

Страница 13: ...module USB stick with software and USB cable Battery operated portable thermal printer only possible together with humimeter USB data interface module described in a separate operating manual Bluetooth module described in a separate operating manual 3 2 2 RH5 1 s scope of supply humimeter RH5 4 pieces of AA Alkaline batteries Operating manual Optional accessories Rubber protection cover Sword sens...

Страница 14: ...positive terminals matching those indicated on the battery compartment Press down the batteries so that they lay flat on the bot tom of the housing figure 4 As soon as all batteries have been inserted the device switches on automatically 4 Push the battery cover onto the housing until it clicks into place Then mount the rubber protection cover onto the housing beginning at the end where the batter...

Страница 15: ...device type are displayed in black and can be selected 2 Use the arrow keys to move from one product type to the next 3 and keep any of them pressed to scroll through the types 4 Confirm your selection by pressing The product type you selected will now be shown at the top of the display 4 3 Taking a measurement For information on how to take a measurement see section 5 The measuring process 4 4 Sw...

Страница 16: ...ble sword sensor holder and the tool for removing the sword sensor holder figure 8 2 Push the sword a few centimetres further into the stack at short intervals approx 10 seconds figure 9 3 Let the device adjust to the material being measu red for an adequate time period see 5 2 Adjust ment behaviour of the sensor 4 Now take the measured values shown on the display of the device Once the reading ha...

Страница 17: ...n adequate time period see 5 2 Adjust ment behaviour of the sensor figure 12 5 Now take the measured values shown on the display of the device Once the reading has been taken it can be saved on the device see 6 2 Saving your readings manu ally or 6 3 Auto save function time based i Information Measuring accuracy This rapid and non destructive measuring procedure allows you to take mois ture readin...

Страница 18: ...rom 20 C to 30 C When using the sword sensor holder pay attention to the temperature adjustment between the sword sensor holder and the sword To demonstrate the importance of temperature adjustment the table below shows the measuring errors due to a temperature difference between the measuring instrument and the material being measured of only 1 C 1 8 F at different ambient tempera tures 10 C 50 F...

Страница 19: ...to be switched on and be in the product selection menu 1 Press twice or hold for 2 seconds 2 Select Options To do so press or and confirm by pressing 3 Select Log Time figure 13 To do so press or and confirm by pressing 4 Select Hold figure 14 To do so press or and confirm by pressing The setting has been saved 5 Press to leave the Options menu 6 Press to leave the main menu 6 1 2 Using the Hold f...

Страница 20: ... Device name 8 Relative air humidity average 6 2 1 Saving individual readings The device can be configured in such a way that the device will save a reading every time a button is pressed This option manual save function is the device s default setting Activating the manual save function in the options menu To do so The device has to be switched on and be in the product selection menu 1 Press twic...

Страница 21: ...rocess The display will now appear as shown in figure 18 3 The data you have inputted can be overwritten at any time 4 Inputting letters Press and hold to quickly scroll to the required letter and either press it for 3 seconds or press to confirm the selected letter figure 19 5 Inputting numbers Press and hold to quickly scroll to the required number and either press it for 3 seconds or press to c...

Страница 22: ...e for the saved reading and to finish the measuring process The display will now appear as shown in figure 21 4 The data you have inputted can be overwritten at any time 5 Inputting letters Press and hold to quickly scroll to the re quired letter and either press it for 3 seconds or press to confirm the selected letter figure 22 6 Inputting numbers Press and hold to quickly scroll to the required ...

Страница 23: ...ired time interval figure 25 To do so press or and confirm by pressing The setting has been saved 5 Press to leave the Options menu 6 Press to leave the main menu 6 3 2 Auto save function Saving measured values To do so The device has to be in the Data Log menu see Data Log menu page 5 1 Press The device will save a reading at the selected time interval The number of data saves will increase by on...

Страница 24: ...it for 3 seconds or press to confirm the selected number 6 Moving forward back Press to switch to another input level Press or to move forward or back 7 Confirm your entry by pressing The data you entered has been saved 6 4 Viewing individual readings To do so You must have saved a reading e g 1 log The display will now appear as shown in figure 28 1 Press 2 Select the required reading To do so pr...

Страница 25: ...n in figure 32 4 Press again The display will now appear as shown in figure 33 5 Navigate to the required reading No 1 No 2 No 3 To do so press 6 Press to leave this screen 6 6 Deleting all measured values data log To do so You must have taken and saved one or several readings 1 Press twice or hold for 2 seconds 2 Select Edit Logs figure 34 To do so press or and confirm by pressing 3 Select Clear ...

Страница 26: ...g or a series of measurements e g 3 logs The display will now appear as shown in figure 37 1 Press 2 Select the required reading To do so press or The display will now appear as shown in figure 38 3 Press to switch to another input level The display will now appear as shown in figure 39 4 Press The display will then show the message clear figure 40 5 Confirm by pressing The value has been deleted ...

Страница 27: ...he display will now appear as shown in figure 43 4 Press 5 The display will now appear as shown in figure 44 6 Select the required measured value To do so press or 7 Press to switch to another input level The display will now appear as shown in figure 45 8 Press to delete the value shown The display will then show the message clear figure 46 9 Confirm by pressing The value has been deleted Deleted...

Страница 28: ...is totally saturated with water vapour If the air has more than 100 humidity the excessive humidity would condense or precipitate as mist Dew Point The dew point is the temperature to which the air that is not completely saturated with water vapour must be cooled so that it is completely saturated When a room with the current relative humidity cools down to the dew point temperature the water vapo...

Страница 29: ... 47 Before using LogMemorizer please refer to the the separate LogMemorizer operating manual for the correct configuration of the USB COM Port For more information on LogMemorizer please refer to the separate LogMemorizer operating manual supplied with the device 8 2 Exporting measured values to a computer To do so LogMemorizer must be installed And you must have taken and saved one or several moi...

Страница 30: ...s or and confirm by pressing 7 Select Manual Logs or Auto Logs figure 50 To do so press or and confirm by pressing The display will then show the message Send figure 51 All of the measuring values saved on the humime ter RH5 will now be sent to your computer Initiating the data export at your computer Connect the humimeter RH5 to your computer using the supplied USB cable 1 Insert the USB Mini B c...

Страница 31: ...cent manually saved logs Import all auto save logs for importing all auto save readings Import most recent auto save log for importing the most recent auto save logs No Name 1 Import all auto save logs 2 Import most recent auto save series 3 Import all manual logs 4 Import most recent manual log The measuring values saved on the humimeter RH5 will now be sent to your computer 54 3 4 2 1 ...

Страница 32: ...formation figure 55 No Name 1 Serial number 2 Software version 3 Battery status 4 Memory status 3 Confirm by pressing 4 Press to leave the main menu 10 Configuring the device 10 1 Turning on Bluetooth The information on Bluetooth is provided in a separate operating manual 10 2 Adjusting the date time 1 Press twice or hold for 2 seconds 2 Select Options To do so press or and confirm by pressing 3 S...

Страница 33: ... 6 Moving back Press to switch to another input level To move backward between DD MM YY and hh mm ss press 7 Confirm the date time by pressing The settings have been saved 8 Press to leave the Options menu 9 Press to leave the main menu 10 3 Selecting a language 1 Press twice or hold for 2 seconds 2 Select Options To do so press or and confirm by pressing 3 Select Language To do so press or and co...

Страница 34: ... four digit password is the device s serial number 4 Inputting numbers Press and hold to quickly scroll to the required number and either press it for 3 seconds or press to confirm the selected number figure 59 5 Moving back Press to switch to another input level To move back press 6 Confirm the four digit password by pressing The setting has been saved The C F BL On Time Auto OFF Time Calibrate M...

Страница 35: ...or Fahrenheit F To do so press or and confirm by pressing The setting has been saved 5 Press to leave the Options menu 6 Press to leave the main menu 10 7 Reducing the device s power consumption 10 7 1 Configuring the display illumination time To do so All of the options must be activated see 10 4 Activating options 1 Press twice or hold for 2 seconds 2 Select Options To do so press or and confirm...

Страница 36: ...the main menu 10 8 Calibrating the device For information on how to calibrate your device see section 12 2 point optionally 3 point calibration 10 9 Configuring the material calibration function The type calibration function is described in a separate operating manual 10 10 Changing the password To do so All of the options must be activated see 10 4 Activating options 1 Press twice or hold for 2 s...

Страница 37: ...ust be activated see 10 4 Activating options 1 Press twice or hold for 2 seconds 2 Select Options To do so press or and confirm by pressing 3 Select Reset To do so press or and confirm by pressing The display will then show the message Reset figure 60 4 Confirm by pressing The device will now be reset to its factory settings All of your personal settings will be lost The display will show the stat...

Страница 38: ...you are responsible by law for pro perly disposing of all used batteries which must not be disposed of as domestic waste Battery Directive 11 2 Care instructions Do not leave the device out in the rain The device is not waterproof Do not immerse the sensor in water Do not expose the device to extreme temperatures Protect the device from strong mechanical shocks and loads 11 3 Cleaning the device A...

Страница 39: ...ation equipment figure 66 3 Insert the second sealing ring 2 figure 64 into the upper part of the calibration equipment figure 67 4 Put the textile pad in the lower part figure 68 and carefully pour the humidity standard onto the pad beginning with the humidity standard of 35 relative humidity 5 Put the third sealing ring 4 figure 64 onto the lower part figure 69 then put the alumi nium disc 3 ont...

Страница 40: ...ture 3 At ideal temperature conditions device calibration equipment and humidity standard have a temperature of 23 C the value printed on the humidity stan dard can be used as reference value 4 In case of deviation from the factory temperature 23 0 C the real humidity value must first be determined according to the table below Temperature Humidity standards 35 50 80 20 C 34 6 49 8 79 9 21 C 34 8 4...

Страница 41: ...for 2 seconds 2 Select Options To do so press or and confirm by pressing 3 Select Calibrate To do so press or and confirm by pressing 4 Select Humidity To do so press or and confirm by pressing The display will then show the first index Index1 Idx 1 figure 71 5 Enter the indices 1 to 6 of the Ic values already sto red in the device into the corresponding fields of the MS Excel calculation sheet pa...

Страница 42: ... to this point no changes have been made to the calibration 10 Now remove the calibration equipment from the sword probe and repeat the procedure from 12 1 Assembling the calibration equipment optionally with the humidity standard 50 relative humidity or the humidity standard 80 relative humidity 12 3 2 Entering the calculated calibration values into the RH5 or RH5 1 To do so The Microsoft Excel c...

Страница 43: ...epeat steps 5 and 6 for the rest of the calculated indices leaving index7 Idx 7 untouched 8 Press to switch to another input level and to leave the Humidity menu The recalibration has been saved 9 Press to leave the Calibrate menu 10 Press to leave the Options menu 11 Press to leave the main menu 12 In case of an incorrect recalibration it is possible to reset the device to its factory settings se...

Страница 44: ... type Check whether you have selected the right product type product before taking a reading see 7 Product types Dripping water or sprayed water Direct contact of the sensor with dripping or sprayed water will destroy it Irreversible damage of the sen sor due to aggressive gases Please contact your dealer Condensation caused by a change in temperature Condensation on the sensor interferes with the...

Страница 45: ...o 60 C 14 2 Disposing of the device Devices marked with this symbol are subject to Directive 2012 19 EU of the European Parliament and of the Council of 4 July 2012 on waste electrical and electronic equipment WEEE If the device is being operated outside the European Union the national regulations on the disposal of such devices that apply in the country of use must be observed Electronic devices ...

Страница 46: ...the applicable requirements of the following documents EN 61326 1 2013 Electrical equipment for measurement control and laboratory use EMC requirements EN 50581 2012 Technical documentation for the assessment of electrical and elec tronic products with respect to the restriction of hazardous substances I hereby declare that the equipment named above has been designed to comply with the relevant Se...

Страница 47: ...5 F bei 77 C Operating temperature 10 C to 60 C Storage temperature 20 C to 60 C Temperature compensation Automatic Data memory Up to 10 000 measuring values Power supply 4 pcs of 1 5 Volt AA Alkaline batteries Current consumption 60 mA incl display illumination Menu languages German English French Italian Spanish Por tuguese Czech Polish Russian International Display 128 x 64 illuminated matrix d...