humimeter RH5 & RH5.1 Operating Manual
Page 42
»
In that case navigate back to Index1 (Idx:[ 1]) by pressing
and press
shorter.
8. Now enter the real humidity value, the temperature, the displayed humidity val-
ue and the new Ic value (point 7) of the humidity standard 35% relative humidity
into the corresponding fields in the MS Excel calculation sheet (figure
»
If you forget to enter one value, the calculation is wrong.
9. Now do not press any other key on the device and wait until it switches itself off
(by default this takes about 5 minutes).
»
Up to this point, no changes have been made to the calibration.
10. Now remove the calibration equipment from the sword probe and repeat the
procedure from
"12.1 Assembling the calibration equipment",
optionally with the
humidity standard 50 % relative humidity or the humidity standard 80 % relative
humidity.
12.3.2 Entering the calculated calibration values into the RH5 or RH5.1
To do so:
The Microsoft Excel calculation sheet has been filled in correctly.
1. Press
twice or hold for 2 seconds.
2. Select
Options
. To do so, press
or
and confirm by pressing
.
3. Select
Calibrate
. To do so, press
or
and confirm by pressing
.
4. Select
Humidity
. To do so, press
or
and confirm by pressing
.
»
The display will then show the first index, Index1
(Idx:[ 1]) (figure
).
74