 Neo Vacuum Module Manual 

• 

Read, understand, and practice the precautionary and operating instructions. 
Know the limitations and hazards associated with using any electrical 
stimulation. Observe the precautionary and operational decals placed on the 
unit.

• 

DO NOT operate this unit in an environment where other devices are being used 
that intentionally radiate electromagnetic energy in an unshielded manner.

• 

All modalities should be routinely checked before each use to determine that all 
controls function normally, especially that the intensity control does properly 
adjust the intensity of the ultrasonic power output in a stable manner. Also, 
determine that the treatment time control does actually terminate ultrasonic 
power output when the timer reaches zero.

• 

Device is designed to comply with electromagnetic safety standards.  This 
equipment generates, uses, and can radiate radio frequency energy and, if not 
installed and used in accordance with instructions, may cause harmful 
interference to other devices in the vicinity.  However, there is no guarantee that 
interference will not occur in a particular installation. Harmful interference to 
other devices can be determined by turning this equipment on and off. Try to 
correct the interference using one or more of the following:

-  Reorient or relocate the receiving device
-  Increase the separation between the equipment
-  Connect the equipment to an outlet on a different circuit from that 

the other device(s) are connected and consult the factory field service 
technician for help.

-  Consult your authorized DJO dealer for help.

• 

Do not operate this unit when connected to any unit other than DJO devices or 
accessories specifically described in user or service manuals.

• 

DO NOT disassemble, modify, or remodel the unit or accessories. This may cause 
unit damage, malfunction, electrical shock, fire, or personal injury.

• 

Failure to use and maintain the Intelect® Neo System, its modules, and its 
accessories in accordance with the instructions outlined in this manual will 
invalidate the warranty.

CAUTION

WARNING

6.  Reattach the Back Panel on the Neo Head Unit.  Secure 

the back panel to the chassis with the two (2) screws 

that were removed in Step 3. 

7.    Reattach the handle and secure with the four (4) 

screws that were removed in Step 3.

8.   Return faceplates to both sides of the Neo Head.  Refer 

to Intelect Neo Clinical Therapy System User Manual, 

item #13-7651, for instructions on installing faceplates.

9.  Replace the Drawer.

10.  Reconnect the power cord to the rear of the device.

MODULE INSTALLATION (CONTINUED)

• 

DO NOT permit foreign materials, liquids or cleaning agents to enter the unit, 
including, but not limited to, inflammables, water, and metallic objects from 
entering the unit, to prevent unit damage, malfunction, electrical shock, fire, or 
personal injury.

• 

Use of parts or materials other than DJO’s can degrade minimum safety.

• 

The Intelect® Neo Clinical Therapy System is not designed to prevent the ingress 
of water or liquids. Ingress of water or liquids could cause malfunction of internal 
components of the system and therefore create a risk of injury to the patient.

• 

The Intelect® Neo Vacuum Electrode Module is designed to operate only when 
properly installed in the Intelect® Neo Clinical Therapy System Cart.

• 

Before each use, inspect Vacuum Electrode Cups and Lead Hoses for cracks and 
damage which may not allow the vacuum to properly secure the electrodes.

• 

Drain the Vacuum Electrode Module water reservoir regularly to prevent 
excessive accumulation from electrode sponge water.

• 

Periodic flushing of the system and reservoir are required to maintain Factory 
functionality of the Vacuum Electrode Module. Refer to the Maintenance Section 
of this manual for proper instructions.

• 

Be sure to read all instructions for operation before treating patient.

• 

Make certain the unit is electrically grounded by connecting only to a grounded 
electrical service receptacle conforming to the applicable national and local 
electrical codes.

• 

Care must be taken when operating this equipment around other equipment. 
Potential electromagnetic or other interference could occur to this or to the 
other equipment. Try to minimize this interference by not using other 
equipment in conjunction with it.

• 

This device should be kept out of the reach of children.

• 

Use of accessories other than those specified in this User Manual may increase 
electrical emissions and decrease electrical immunity of the device.

• 

DO NOT operate this unit in an environment where other devices are being used 
that intentionally radiate electromagnetic energy in an unshielded manner.

• 

Contaminated sponges, electrodes, and leadhoses can lead to infection.

• 

Use of contaminated sponge or electrodes with corrosion with vacuum system 
can result in possible infection or skin irritation.

• 

Clean vacuum sponges, electrodes, and leadhoses before each use otherwise it 
can lead to cross-contamination and infection.

• 

Stop treatment immediately if patient experiences discomfort or pain.

Содержание Neo Vacuum Module

Страница 1: ...oglobal eu fr_FR index html 2017 DJO LLC All rights reserved NeoVacuum Module Manual 13 8809 Rev E For complete User Operating Instructions including Cautions Warnings Dangers Indications and Contraindications refer to the Intelect Neo ClinicalTherapy System User Manual Item 13 7651 CDVersion item 13 7652 To obtain a copy contact your local representative or DJO Customer Care 0086 ...

Страница 2: ... 70006 Vacuum hoses 1 2 Vacuum hoses 3 4 60mm electrodes 60mm Sponges Tools required not included 2 Phillips screwdriver The System is programmed to automatically recognize the new Module therefore no software installation or configuration is required Complete the following steps to install the module 1 Power off device Remove the power cord from the rear of the device 2 Remove the back panel from...

Страница 3: ...ach side of the handle Remove the back panel from the Neo Head 4 Insert the Vacuum Module into the Cart as shown Attach the Vacuum Module with the four 2 Phillips screws removed from the back panel in Step 2 5 Thread the Ribbon Cable up to the receptacle on the circuit board plug in and secure with locking clips as shown If needed remove the cable between the power supply and the board prior to se...

Страница 4: ... Intelect Neo Clinical Therapy System User Manual item 13 7651 for instructions on installing faceplates 9 Replace the Drawer 10 Reconnect the power cord to the rear of the device MODULE INSTALLATION CONTINUED DONOTpermitforeignmaterials liquidsorcleaningagentstoentertheunit including butnotlimitedto inflammables water andmetallicobjectsfrom enteringtheunit topreventunitdamage malfunction electric...

Страница 5: ...ntenance installation removalorreplacementprocedurestoprevent electricalshockandpossibledamagetosystem Vacuumelectrodesshouldnotbeusedonpatientswiththin paperyskin Vacuummayleadtocontactdifficultyandbruising Vacuumelectrodesarenotsuitableforpatientswhoaretakingsteroids dueto thelikelihoodofbruising Handle cleananddisposeofcomponentsandaccessoriesthat havecomeincontactwithbodilyfluidsaccordingtoNat...

Отзывы: