ELECTROMAGNETIC COMPATIBILITY (EMC) TABLES (CONTINUED)
Recommended separation distances between portable and mobile RF communications equipment and the Intelect® Mobile 2
The Intelect® Mobile 2 is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the Intelect®
Mobile 2 can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and
the Intelect® Mobile 2 as recommended below, according to the maximum output power of the communications equipment.
Rated maximum output
power of transmitter
W
Separation distance according to frequency of transmitter
d (m)
150 kHz to 80 MHz
outside ISM bands
D = 1.2 √P
150 kHz to 80 MHz
in ISM bands
D = 2 √P
80 MHz to 800 MHz
D = 1.2 √P
800 MHz to 2,5 GHz
D = 2.3 √P
0.01
0.12
0.20
0.12
0.23
0.1
0.38
0.63
0.38
0.73
1
1.2
2.0
1.2
2.3
10
3.8
6.3
3.8
7.3
100
12
20
12
23
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in metres (m) can be determined using the equation applicable
to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 The ISM (industrial, scientific and medical) bands between 150 kHz and 80 MHz are 6,765 MHz to 6,795 MHz; 13,553 MHz to 13,567 MHz;
26,957 MHz to 27,283 MHz; and 40,66 MHz to 40,70 MHz.
NOTE 3 An additional factor of 10/3 has been incorporated into the formulae used in calculating the recommended separation distance for
transmitters in the ISM frequency bands between 150 kHz and 80 MHz and in the frequency range 80 MHz to 2,5 GHz to decrease the likelihood
that mobile/portable communications equipment could cause interference if it is inadvertently brought into patient areas.
NOTE 4 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures,
objects and people.