2 1
PATIENT PREPARATION
E N
INTELECT
®
MOBILE 2 ULTRASOUND USER MANUAL
ULTRASOUND PATIENT PREPARATION
1.
Examine the skin for any wounds and clean the skin
2.
View the Applicator recommendation in the treatment
guidelines.
3.
Review guidelines for Ultrasound (as a reference
point only) on the treatment review screen prior to
administering treatment.
NOTE: Applicators are available in the sizes shown below:
1
2
3
4
Applicator Preparation and Use
1.
Clean applicator before each therapy session with
warm soapy water, check the applicator has no cracks
prior to use.
2.
Liberally apply transmission gel to the treatment area
on the patient.
3.
Move the applicator during therapy session in a circular
motion. The area treated should be:
• Twice the diameter of the applicator
• For 5cm2 US applicator: three times the diameter
of the applicator if output power > 4 W, Continuous
mode.
The applicator should always be held by the grip and not by
the Ultrasound Applicator head.
4. If US Coupling is “On”, the Applicator is properly coupled
to the patient and administering ultrasound when the
LED is constantly illuminated. If the applicator head
becomes uncoupled the LED on the head will flash. If
"US coupling" setting is ON, several beeps will be also
heard until the head is coupled again. Treatment time
stops during uncoupling.
NOTE: Ultrasound output will continue to be emitted in all US
coupling modes even if the applicator is uncoupled. The output power
is reduced to a very low level to prevent ultrasound head warming.
For ULTRASOUND OPERATION, refer to page 60