GENERAL WARNINGS AND PRECAUTIONS
E N
INTELECT
®
MOBILE 2 ULTRASOUND USER MANUAL
WARNING
• This device should be used only under the continued supervision of a physician or licensed practitioner.
• Contaminated sponges, electrodes, leadwires, and gel can lead to infection.
• DO NOT operate the Intelect® Mobile 2 within the vicinity or environment of an ultrasonic diathermy system.
• DO NOT operate the Intelect® Mobile 2 within the vicinity or environment of any microware and RF shortwave
diathermy system.
• DO NOT operate this unit in an environment where other devices are being used that intentionally radiate
electromagnetic energy in an unshielded manner.
• Portable RF communications equipment should be used no closer than 30 cm (12 inches) to any part of the Intelect
Mobile 2, including cables specified by the manufacturer. Otherwise, degradation of the performance of this equipment
could result.
• Use of accessories, transducers and cables other than those specified or provided by the manufacturer of this equipment
could result in increased electromagnetic emissions or decreased electromagnetic immunity of this equipment and result
in improper operation.
• Battery replacement by inadequately trained personnel could result in fire or explosion. Please read carefully the battery
replacement instructions in the Mobile 2 IFU before attempting to replace the battery.
• Device is designed to comply with electromagnetic safety standards. This equipment generates, uses, and can radiate
radio frequency energy and, if not installed and used in accordance with instructions for use, may cause harmful
interference to other devices in the vicinity. Harmful interference to other devices can be determined by turning this
equipment on and off. Try to correct the interference using one or more of the following:
»
Reorient or relocate the receiving device
»
Increase the separation between the equipment
»
Connect the equipment to an outlet on a different circuit from that to which the other device(s) are connected
»
Consult your authorized DJO dealer for help.
• Disconnect the system from the power source before attempting any maintenance, installation, removal, or replacement
procedures to prevent electrical shock and possible damage to system.
• The Intelect® Mobile 2 may be susceptible to Electro-Static Discharge (ESD) at greater than ±6 kV when first grasping
the Ultrasound applicator. In the event of such a discharge, the Intelect® Mobile 2 may display a permanent error. The
Intelect® Mobile 2 will terminate all active outputs (stim, ultrasound,), automatically place the unit in a safe state.
• To prevent Electro-Static Discharge (ESD) at greater than ±6 kV:
»
Grasp and hold the Ultrasound prior to starting treatment. If the applicator must be put down prior to completion
of treatment, stop the current treatment first and then place the applicator in the holder.
»
Maintain humidity in the use environment to at least 50% relative humidity.
»
Floors should be wood, concrete, or ceramic tile. If floors are covered with synthetic material, DJO recommends
implementing additional controls to maintain relative humidity to at least 50%.
»
Communicate these ESD-precautionary procedures to healthcare staff, contractors, visitors, and patients.