3
Read, understand, and practice the precautionary and operating
instructions. Know the limitations and hazards associated with
using any electrical traction device. Observe the precautionary and
operational decals placed on the unit.
Use this unit only with tables, stands, power cord, and accessories
specifically designed and manufactured by Chattanooga Group.
Not adhering to this guideline may cause damage to the unit,
malfunction, electrical shock, fire, or personal injury.
This unit should be operated, transported, and stored in
temperatures between 60° F and 104° F (16° C and 38° C) or
atmospheric pressure range between 700-1060 hPa, with relative
humidity ranging from 30% - 60%.
Do not operate the unit when connected to any unit other than
Chattanooga Group devices. Do not use devices manufactured by
other companies on Chattanooga Group equipment. Chattanooga
Group is not responsible for any consequence resulting from using
products manufactured by other companies.
The unit should be routinely checked before each use to determine
all controls function normally.
Handle the unit with care. Inappropriate handling of the unit may
adversely affect its characteristics.
Before each use, inspect the Traction Cord for wear. Prolonged wear
on the cord will cause it to break, which may cause sudden release of
traction pressure on a patient.
Test the Patient Interrupt Switch cable before each use for proper
operation.
Do not use the Clevis as a handle to pick up or carry the unit.
•
•
•
•
•
•
•
•
•
Do not remove the cover. This may cause unit damage, malfunction,
electrical shock, fire, or personal injury. There are no user-serviceable
parts inside the unit. If a malfunction occurs, discontinue use
immediately, disconnect the Mains Power Cord from the outlet, and
consult the dealer for repair service.
Do not use sharp objects such as a pencil point or ballpoint pen to
operate the buttons on the Touch Screen as damage may result.
Do not permit any foreign materials or liquids to enter the unit. Take
care to prevent any foreign materials including, but not limited to,
inflammables, water, and metallic objects from entering the unit.
These may cause unit damage, malfunction, electrical shock, fire, or
personal injury.
Do not disassemble, modify, or remodel the unit or accessories.
This may cause unit damage, malfunction, electrical shock, fire, or
personal injury.
Do not use the unit near devices such as X-ray units or diathermy
units. These units may emit high frequency noise that may affect the
operation of the unit.
If you have difficulty operating the unit after carefully reviewing
this operator’s guide, contact your Chattanooga Group dealer for
assistance.
Failure to use and maintain the traction unit and its accessories
in accordance with the instructions outlined in this manual will
invalidate your warranty.
The Patient Interrupt Switch is essential for the functioning of the unit.
If it is not connected, or if it is malfunctioning, the unit will not work.
•
•
•
•
•
•
•
•
ABOUT TRACTION THERAPY