SETUP
MOUNTING UNIT ON TABLE
The TX traction unit was designed to be utilized with Chattanooga
Group traction tables.
NOTE:
The TX traction unit should be mounted facing the direction of
pull so the Traction Cord does not contact the side of the slot
from which it extends.
1. To secure the unit to the
mounting surface, push the Lock
Knob down and turn left to
loosen the clamp. Fit the unit to
the mounting surface and turn
right to tighten the clamp down
onto the mounting surface. Push
the Lock Knob again to secure the
unit to the mounting surface.
2. Remove the cover on the back
of the unit and connect the Mains
Power Cord (female end) to the
Mains Power Cord Receptacle.
3. Verify that there is a good
connection between the Mains
Power Cord and the Mains Power
Cord Receptacle. Always
make certain that the Mains
Power Cord is properly plugged
into the unit.
4. Check the voltage rating on the
serial plate located on the back of
the unit. Plug the Mains Power
Cord into a 100 V-24 V AC outlet,
as required. Replace the cover on
the back of the unit.
• To prevent accidental disengagement, this unit must be securely
attached to the mounting surface of the pedestal or traction stand. It
is the responsibility of the user to verify the adequacy of the
installation before use in patient therapy.
• Do not operate the unit when connected to any unit other than
Chattanooga Group devices. Do not use devices manufactured by
other companies on Chattanooga Group equipment. Chattanooga
Group is not responsible for any consequence resulting from using
products manufactured by other companies.