Quick Start Guide for 3680A/B Cable and Antenna Analyzer
Specific mechanical calibration procedure is as follows:
Connect the open to the test port and select the [OPEN] menu. At this time, the screen will
automatically prompt “[OPEN]Measuring
……
”. When the [Open] menu becomes into [OPEN], it
indicates the completion of the open circuit calibration. Figure 3.3 below indicates the completion of
open circuit measurement.
Figure 3.3 Completion of Open Calibration
Connect the short to the test port and select the [Short] menu. At this time, the screen will
automatically prompt “[SHORT]Measuring
… …
”. When the [SHORT] menu becomes into
[SHORT], it indicates completion of the short circuit calibration.
Connect the load to the test port. Then, select the [Load] menu, at this time, the screen will
automatically prompt “[LOAD]Measuring
……
”. When the [Load] menu becomes into [Load], it
indicates completion of the load calibration, as shown in Figure 3.4. At this moment, select the
[Done] menu to finish mechanical calibration process.