32
CETONI Nemesys S Hardware Manual
5.12
Universal Syringe Holder Nemesys S
Before mounting a syringe on the CETONI Nemesys S, it must be configured and selected in the software.
The procedure is described in the software manual. For this you need the Volume (
Scale Volume
), the
nominal stroke (
Scale Length
) and the maximum stroke (
Piston Stroke
). In combination with the counter
holder, you can use syringes that are between 89 and 119 mm long when empty
excluding the connection nozzle and have a diameter between 8 and 30 mm. The
operation is explained below:
(1)
Push the syringe piston completely into the syringe and bring the piston holder to the foremost
position (syringe empty) via the software.
(2)
Unscrew the clamping screw
1
of the syringe down holder
2
completely and rotate the syringe down
holder to the side or pull it upwards off the guide rod
3
so that you can insert the syringe. (To adjust
the height, reach as close as possible to the guide rod).
(3)
Loosen the clamping screw
4
of the piston holder and push the movable part
5
forward. If necessary,
select an adapter plate
6
matching your syringe piston from the storage position
7
.
(4)
Loosen the clamping screw
8
of the counter holder
9
and pull it out so far that you can insert the
syringe.
1
3
2
4
6
7
8
9
5