24
CETONI Nemesys S Hardware Manual
5.5
Fluidics
Two syringe pumps equipped with Contiflow valves are required for uninterrupted fluid dispensing. A
Contiflow valve has three connections with ¼"-28 UNF thread. The connection for the syringe is on one
side and is not labeled. The syringe is screwed directly into this connection.
The connections for application and reservoir are on the opposite side and are labeled RES and APP. For
continuous operation, connect the APP connections of both devices, for example via a T-piece, with your
application (blue in the picture) and both RES connections to your reservoir (orange in the picture).
An LED under the respective port lights up to indicate the current switching position. If the LED lights
green, then this port is connected to the syringe. Tube to the reservoir.
For information on how to configure the Continuous Flow in CETONI Elements, please refer to the
software manual.
ATTENTION
. Only use fittings and hoses that are approved for the expected
pressure and check the tightness of all fluidic connections after connection and at
regular intervals.
LEDs
Tube to the reservoir
Tube to the application