CETONI Nemesys S Hardware Manual
25
IMPORTANT
.
Some valve types have a hand lever for manual actuation. Please leave
this in the normal position shown. Otherwise, the proper functioning of the Contiflow
Valve cannot be guaranteed.
5.6
Upright Operation
For easier bleeding of the syringes, it is often helpful to operate the syringe pump upright. To increase
the standing safety in this case, the CETONI Nemesys S is equipped with a stand foot. This is magnetically
located in a recess on the side. Take it out, turn the module upright and put the stand foot back in place
rotated by 90°.
ATTENTION
. Place the devices at least 30 cm from the edge of the table, even when
using the stand foot. For permanent upright operation of larger systems, we
recommend using the optional system clamp.
Remove Stand Foot
Insert Stand Foot
Hand lever