background image

44

2. Attached documents

The CGM MB-1101 is intended for use in the electromagnetic environment

specified below. The customer or the user of CGM MB-1101 should assure

that it is used is such an environment.

Immunity test

Compliance level

IEC 60601

Test level

Electromagnetic

environment - guidance

Electrostatic

discharge

(ESD)

IEC 61000-4-2

6kV Contact

8kV Air

6kV Contact

8kV Air

Floors should be wood,

concrete or ceramic tile.

If floors are covered with

synthetic material, the

relative humidity should be

at least 30%. 

Electrical fast

Transient/burst

IEC 61000-4-4

2kV for power

supply lines 1kV

for input/output

lines

2kV for power

supply lines 1kV

for input/output

lines

The main power’s quality 

should be that of a typical

commercial or hospital

environment.

Surge

IEC 61000-4-5

1kV differential

mode

2kV common

mode

1kV differential

mode

2kV common

mode

The main power’s quality 

should be that of a typical

commercial or hospital

environment.

Power frequency

(50/60Hz)

Magnetic field

IEC 61000-4-8

Voltage dips, 

and Voltage

variations on

power supply

input line

IEC 61000-4-11

3.0 A/m

3.0 A/m

Power frequency magnetic

fields should be at levels

characteristic of a typical

location in a typical

commercial or hospital

environment.

Note : 

U

<5% 

U

(>95%

dip in 

U

for 0.5 cycle

40% 

U

in 

U

70% 

U

in 

U

<5% 

U

(<95%

dip in 

U

for 5 s

<5% 

U

(>95%

dip in 

U

for 0.5 cycle

40% 

U

in 

U

70% 

U

in 

U

<5% 

U

(<95%

dip in 

U

for 5 s

The main power’s quality 

should be that of a typical

commercial or hospital

environment. If the user of

the CGM MB-1101 requires

continued operation during

power mains interruptions,

it is recommended that the

CGM MB-1101 

be powered

from an uninterruptible

power supply or a battery.

Manufacturer’s declaration - electromagnetic immunity

45

2. Attached documents

The 

  is intended for use in the electromagnetic environment specified 

below. 

The customer or the user of the 

  should assure that it is used in such an environment

 

Immunity 

test 

IEC 60601 

Test level 

Compliance 

level 

Electromagnetic environment -

guidance 

Conducted 
RF 
IEC 61000-4-
6

 

3 Vrms 

150 kHz to 80 
MHz

 

3 Vrms 

150 kHz to 80 
MHz

 

Portable and mobile RF communications 
equipment should be used no closer to any part 
of the 

CERAGEM-CH

, including cables, than 

the recommended separation distance calculated 
from the equation applicable to the frequency of 
the transmitter. 
 

Recommended separation distance 

 

 

Radiated RF 
IEC 61000-4-

 

3 V/m 

80.0 MHz to 2.5 
GHz 

3 V/m 

80.0 MHz to 2.5 
GHz 

Recommended separation 

distance 

 

 

 

Where 

is the maximum output power rating of 

the transmitter in watts (W) according to the 
transmitter manufacturer and 

is the 

recommended separation distance in meters (m). 
Field strengths from fixed RF transmitters, as 
deter-mined by an electromagnetic site survey,   
(a) Should be less than the compliance level in 
each frequency range (b). 

 

Interference may occur in the vicinity of 
equipment marked with the following symbol:

 

 

Note 1)

 

U

 is the A.C. mains voltage prior to application of the test level. 

 

Note 2)

 At 80 MHz and 800 MHz, the higher frequency range applies. 

 

Note 3)

 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and 

reflection from structures, objects and people.

 

Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile 

radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. 
To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be 
considered. If the measured field strength in the location in which the EUT is used exceeds the applicable RF 
compliance level above, the EUT should be observed to verifynormal operation. If abnormal performance is observed, 
additional measures may be necessary, such as re-orienting or relocating the EUT. 
 

b

 Over the frequency range 150 kHz to 80 MHz, field strengths should be less than [V1] V / m.

 

CGM MB-1101

CGM MB-1101

CGM MB-1101

80 MHz to 800 MHz

800 MHz to 2.5 GHz

Содержание CGM MB-1101

Страница 1: ... Manual http www ceragem net TEL FAX 82 41 523 2560 82 41 523 2550 3699 Wilshire Blvd Ste 930 Los Angele CA 900010 US TEL 213 480 7070 FAX 213 480 7071 http www ceragem net CERAGEM CO LTD CERAGEM CO LTD US ...

Страница 2: ...of violating the instructions there is an emergent possibility of serious damage or even death Warning In case of violating the instructions there is some possibility of serious injury or death Caution In case of violating the instructions there is possibility of minor injury or property damage The information below stands to protect the safety of the user and to prevent any property damage All us...

Страница 3: ...all the customer center when in need of a repair There is a risk of electrick shock fire injury and burns 15 Please take care not to insert your hand or any other body part into the main equipment or auxiliary parts There is a risk of electric shock and injury 16 Device is only intended for use by people age 18 and over 17 To avoid the risk of injury If you have a pacemaker defibrillator other art...

Страница 4: ...vice must accept any interference received including interference that may cause undesired operation NOTE This equipment has been tested and found to comply with the limit for a Class B digital device pursuant to Part 15 of the FCC Rules These limits are designed to provide reasonable protection against harmful interference in a residential installation This equ ipment generates uses and can radia...

Страница 5: ...6 MEMO 1 Product s characteristics 2 The name of each part of the product 7 ...

Страница 6: ...lim and light remote control has function keys in the form of icons for more convenient product operation and simpler manipulation 4 External projectors 3 sphere 9 sphere provided External projectors that can be used while this massaging equipment is operated are provided as basic items users may select 3 sphere or 9 sphere projectors for the arms and abdomen 5 Indications for Use The intended use...

Страница 7: ...10 MEMO and preparation 1 Check product s composition 2 Check before product installation 3 Installation 4 Check after installation 11 ...

Страница 8: ...e external projector User manual Information on how to use the product SD card Users may add their preferred music in addition to the Ceragem default sound Head cushion This is attached to the top of the product Outer fabric Protects against product contamination and burns this detachable fabric makes washing easy Assembly guide Product assembly methods are described 11 12 1 2 3 4 5 6 7 8 9 10 Rem...

Страница 9: ... cable properly Before connecting the power cord connect the remote control first Connect the power cord onto the main body and plug into a socket If the power cord is connected when the power switch is On it can cause mechanical failure Caution SD CARD Caution Warning Do not touch the connector pin Can cause bodily harm or mechanical failure Do not touch signal input signal output or other connec...

Страница 10: ...uct Caution 1 Before connecting the power cord connect the remote control first 2 According to the user set the temperature appropriately when using the product 3 When sitting down and using the product open the supporting frame completely 4 After using the product turn off the power on the remote control and then turn off the main body ...

Страница 11: ...9 Semi Automatic Manual mode When the mode is in operation the time icon is turned on displaying the operating time by having the blank disappeared one by one 2 3 4 1 5 6 External Internal projector Main Auxiliary heating element Music number Start Pause Volume Power button Mode 1 button Mode 2 button Up Down button Strength button Volume button L P ...

Страница 12: ...tion Check the current strength level Pressing the strength key causes the set point to flash After pressing the strength key adjust the temperature to the desired level using the Up Down key Strength can be adjusted in 1 6 levels Check the changed strength set point on the remote control display 2 How to use the remote control 21 Adjusting Strength The strength of the internal projectors can be a...

Страница 13: ...eck the current condition of sound 2 Press the sound key and select the desired setting using the Up Down key Check the condition of the changed sound on the remote control display 2 How to use the remote control 23 Adjusting Sound Music Selection Start Pause Volume 0 10 Ceragem Sound 1 10 User Sound 11 100 The sound settings can be adjusted while in use ...

Страница 14: ...g the Up Down key 3 Check the changed temperature set points on the remote control display When the temperature setting is completed the temperature is converted into the current level after a few seconds When the auxiliary part is closed the external heating element and the Thermal feature can be used External Projector Auxiliary Heating Element Automatic Mode 3 How to use the product by mode 25 ...

Страница 15: ...duct by mode 26 2 Mode 1 Mode 2 18 minutes Mode 3 Mode 4 Mode 5 Mode 6 Mode 7 Mode 8 Mode 9 3 How to use the product by mode Press twice to select Mode 1 Press twice to select Mode 2 Long press Select modes 3 9 in sequence ...

Страница 16: ... 3 5 7 3 1 4 2 12 10 8 6 4 2 10 seconds Standby 3 5 7 4 12 10 8 6 4 1 3 1 2 2 10 seconds Standby 3 How to use the product by mode 29 Automatic Mode Mode 1 Mode 2 Mode 3 Ceramic moving pattern by mode Thermal massage 88 seconds 2 minutes Standby Abdomen switch 3 5 7 4 12 10 8 6 4 1 3 1 2 2 10 seconds Standby 3 5 7 4 12 10 8 6 4 1 3 1 2 2 10 seconds Standby 60 seconds Standby Abdomen switch Thermal ...

Страница 17: ...ern by mode Thermal massage 88 seconds The massage patterns may be adjusted depending on the user s height Mode 8 Mode 9 3 5 7 4 12 10 8 6 4 3 5 7 4 12 10 8 6 4 10 seconds Standby 1 3 1 2 2 10 seconds Standby 1 3 1 2 2 Ceramic moving pattern by mode 3 How to use the product by mode 31 Thermal massage 88 seconds Semi automatic Mode 3 5 7 4 12 10 8 6 4 1 3 1 2 2 Up Down 3 Times Thermal massage 60 se...

Страница 18: ...d both sides of the main body frame and lift and move horizontally Caution When lifting the equipment please lift slowly while straightening your back and bending your knees Caution Disconnect the power cord remote control and external projector from the main body Then lift the head cushion and close the supporting mat To open the supporting mat pull forward and then lower it Close Open SD C ARD ...

Страница 19: ...ad or skin wipe with a dry cloth 3 If there is contact with cosmetic products or volatile emulsions and is left for a long time the surface may peel off so wipe with a dry cloth as soon as possible 4 When washing the outer cloth use a neutral detergent and wash alone by hand 5 If there is contact with hot objects or heat the leather will curl up so please use caution 6 Avoid direct sunlight or hum...

Страница 20: ...ernal Projector Internal Projector Internal circuit Heating element error Symbol Symbol Error location VEE Vod Overloading motor error Overloading motor error Vertical motor drive Are the external projectors working Q Are the external projectors connected properly A Connect the external projectors again Is the product malfunctioning Q Can you move the internal projectors using the Up Down key A In...

Страница 21: ...Consumer Protection Law If you purchase a product that has Ceragem s approval warranty will last for 1 year and component repairs and services will be provided for free This warranty is only valid in the country that the product has been purchased in Contact the customer center in case of a breakdown This product s component parts will be kept for a term of 3 years after the discontinuance of the ...

Страница 22: ...40 MEMO 2 Attached documents 41 ...

Страница 23: ...ed for domestic purposes Classification Form of electric shock protection Degree of electric shock protection Degree of protection from harmful invasion of water Depending on the mode of operation Continuous operation TYPE BF APPLIED PART Ordinary Do not disassemble and contact the customer service center when problem occurs with a fuse Pull the marked area to exchange the fuse related to service ...

Страница 24: ... an environment Immunity test IEC 60601 Test level Compliance level Electromagnetic environment guidance Conducted RF IEC 61000 4 6 3 Vrms 150 kHz to 80 MHz 3 Vrms 150 kHz to 80 MHz Portable and mobile RF communications equipment should be used no closer to any part of the CERAGEM CH including cables than the recommended separation distance calculated from the equation applicable to the frequency ...

Страница 25: ...below according to the maximum output power of the communications equipment Rated maximum output power W of transmitter Separation distance m according to frequency of transmitter 150 kHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2 5 GHz 0 01 0 12 0 12 0 23 0 1 0 37 0 37 0 74 1 1 17 1 17 2 33 10 3 69 3 69 7 38 100 11 67 11 67 23 33 For transmitters rated at a maximum output power not listed above the...

Отзывы: