42
1.
Explanation of mark, delivery, and storage
43
2. Attached documents
Guidance and Manufacturer’s declaration
- electromagnetic emission
The Model CGM MB-1101(CERAGEM CO., LTD. Automatic Thermal Massager)is
intended for use in the electromagnetic environment specified below.The
customer or the user of CGM MB-1101 should assure that it is used in such an
environment
Emission test
Compliance
Electromagnetic environment - guidance
RF emissions
CISPR 11
Group 1
The Model CGM MB-1101 uses RF energy
only for its internal function. Therefore. Its
RF emissions are very low and are not likely
to cause any interference in nearby electronic
equipment
Harmonics emission
IEC 61000-3-2
Voltage fluctuation
IEC 61000-3-3
N/A
N/A
The Model CGM MB-1101 is suitable for use
in all establishments, including domestic
establishments and those directly connected
to the public low-voltage power supply
network that supplies buildings used for
domestic purposes.
Classification
Form of electric shock protection
Degree of electric shock protection
Degree of protection from harmful invasion of water
Depending on the mode of operation
Continuous operation.
TYPE BF APPLIED PART.
Ordinary
Do not disassemble and contact the customer
service center when problem occurs with a fuse.
* Pull the marked area to exchange the fuse (related to service engineer)
* The equipment has to be installed at place built protective device having appropriate breaking capacity
so that prospect short-circuit current do not exceed 40 A
Capacity of a fuse: AC 250V, T6.3AL
Time-Lag,
6.3A,
Glass Body Fuse
: TYPE BF APPLIED PART
V~ : AC (alternating current) voltage source
: Attention, consult Accompanying documents
: This way up
: Keep dry
: Fragile ; handle with care
: Use no hooks
CERAGEM
The disposal of waste products / residues, etc.
Condition for leakage current value
(Automatic Thermal Massager)
WITH RESPECT TO ELECTRIC SHOCK
FIRE, AND MECHANICAL HAZARDS ONLY
IN ACCORDANCE WITH UL60601-1/CAN/CSA C22.2 No. 601.1
47AB
: General prohibition sign
Condition for Transport and Storage
Condition for Operation