Page 16
STEP 4:
Rotate the probe so that the nearest run-out adjustment set screw is above the finger tip of the indicator.
Using the 3/32” hex wrench slowly turn the setscrew directly above the indicator finger, in the counter clock-wise direction,
the dial indicator will show the reduction in the run-out. Stop loosening this set screw and tighten the set screw that was
nearest the low run-out position.
Do not allow the probe to become loose on the mounting shank by adjusting in small
increments.
Adjust in increments that are about one third of the remaining run-out.
Run Out
0