65
System Specifications
System Specifications
Location
• Designated area with a sturdy, non-flammable workbench capable of handling at least the weight of the system
• SMART 6 & ProFat Weight = 55 lbs (25 kg)
• SMART Q Weight = 38 lbs (18 kg)
• Provides adequate space for sample preparation
• Has a dedicated, grounded outlet no more than 5 ft (1.5 m) from unit
• Free from vibration of large equipment
• Low draft and low dust setting
• Provides clear visual access to the display of the system
• Allows for easy access to the cavity and room for the cover to extend fully
Instrument Dimensions
• 13” (33 cm) W x 23” (58.5 cm) D x 14.5” (36.8 cm) H
• Recommended counter space: 29” (99 cm) W x 29” (99 cm) D x 24” (61 cm) H
• Provides 8” (20 cm) of open space on each side, 6” (15 cm) open space in the rear for ventilation and 9.5”
(24 cm) for overhead clearance when the cover is closed
Electrical Requirements
• SMART Q = 100-240 VAC, 50/60 Hz, 6.3 A
• SMART 6 & ProFat = Region Specific
• 100-200 VAC, 50/60 Hz, 15 A
• 200 - 240 VAC, 50/60 Hz, 10 A
• Line voltage must not vary more than 10% of its specified level
• Dedicated, grounded outlet
• Instrument is supplied with a power cord or an adequately rated power cord may be used
Environmental Constraints
• Indoor use only
• Altitude up to 2000 m
• Operating temperature range of 41°F (5°C) to 104°F (40°C)
• Relative humidity range of 10 % to 85 %, non-condensing
• Overvoltage Category II
• Pollution degree 2
Data Output
• 0.01% resolution
• 0.01% to 99.99% range
Balance
• 25 g capacity
• 0.1 mg resolution