29
Method - Create, Edit and Delete
5.
Select the “Parameters” tab.
5.1.
Once in the system parameters, determine if the stage will be a Set Time or Constant Weight analysis.
NOTE
If creating multiple stages, only the final stage can be a Constant Weight stage. All previous stages must
use Set Time.
5.2.
Set Time - The method will run for a preset length of time, regardless of sample size or final weight.
5.2.1.
Set the Mag Power and IR power . The power should be set at 0%, 25%, 50%, 75%, or 100%
5.2.2.
Set the run time for each analysis.
5.2.3.
Set the Maximum temperature for the sample type. Some category parameters contain a con-
Содержание SMART 6