145
avoid dirt accumulation or deterioration
of the instrument.
Clean
Thermoplastic
Printhead
Remove residue of thermoplastic
granules by heating the printhead with
the cartridge mounted but not the nozzle
and supply compressed air with the BIO
X6.
Clean the printhead body with a smooth
cloth to remove any dust and debris.
When not being used, store the
printhead inside a sealed plastic bag in a
cool and dry place to avoid dirt
accumulation or deterioration of the
instrument.
Every time it
is used
Clean mirrors
(metal reflectors)
The ultrasound sensors are very
sensitive elements that help the BIO X6
self-calibration process. Any type of
residue, speck, dust, or dirt can
considerably affect its performance.
Cover the sensors before you clean the
area around them. Make sure dirt is not
deposited inside the tubes that inside the
metal reflectors and that reflectors are
held firmly place. If there is visible dirt
and the autocalibration function of the
BIO X6 does not work properly, contact
the support team for instruction on how
to clean the system and to check for
proper operation.
Every 3
months
Содержание BIO X6
Страница 1: ...BIO X6 User Manual v 1 0 1 BIO X6 User Manual...
Страница 8: ...BIO X6 User Manual v 1 0 8 01 Safety...
Страница 14: ...BIO X6 User Manual v 1 0 14 02 Specifications...
Страница 19: ...BIO X6 User Manual v 1 0 19 03 Getting started...
Страница 29: ...BIO X6 User Manual v 1 0 29 04 Bioprinting...
Страница 48: ...BIO X6 User Manual v 1 0 48 Figure 22 Custom well plate option...
Страница 68: ...BIO X6 User Manual v 1 0 68 Figure 42 Manual calibration glass slide...
Страница 94: ...BIO X6 User Manual v 1 0 94 05 Gcode Editor...
Страница 104: ...BIO X6 User Manual v 1 0 104 Figure 77 Error message Command not supported...
Страница 105: ...BIO X6 User Manual v 1 0 105 06 Draw Print...
Страница 117: ...BIO X6 User Manual v 1 0 117 07 Shape Generator...
Страница 122: ...BIO X6 User Manual v 1 0 122 08 Print report...
Страница 126: ...BIO X6 User Manual v 1 0 126 Figure 97 Protocol setup continued...
Страница 127: ...BIO X6 User Manual v 1 0 127 Figure 98 Example of data included in print report printhead pressure and print speed...
Страница 130: ...BIO X6 User Manual v 1 0 130 09 Coaxial printing...
Страница 134: ...BIO X6 User Manual v 1 0 134 10 Maintenance...
Страница 146: ...BIO X6 User Manual v 1 0 146 11 Terms and conditions...
Страница 166: ...BIO X6 User Manual v 1 0 166 12 Printhead manuals...
Страница 175: ...BIO X6 User Manual v 1 0 175...
Страница 176: ...BIO X6 User Manual v 1 0 176 Create the future of medicine...