10 APPENDIX A- Technical references
Clause
6.8.3.201 Technical Description - Tab 206 Recommended separation distances
between portable and mobile RF communications equipment and the equipment or
system - for equipment and systems that are not life-supporting
TABLE: Recommended separation distances between portable and mobile RF communications
equipment and the equipment
The device
NewTom 5G
Vet
is intended for use in the electromagnetic environment in which radiated RF
disturbances are controlled. The customer or the user of the device
NewTom 5G Vet
can help prevent
electromagnetic interference by maintaining a minimum distance between portable and mobile RF
communications equipment and the device
NewTom 5G Vet
as recommended below, according to the
maximum output power of the communication equipment.
Rated maximum output power of transmitter
(W)
Separation distance according to frequency of transmitter
(m)
150 kHz to 80 MHz
d=
80 MHz to 800 MHz
d=
800 MHz to 2,5 GHz
d=
0,001
0.037
0.037
0.072
0,1
0.37
0.37
0.72
1
1.2
1.2
2.3
10
37.9
37.9
7.27
100
120
120
23
For transmitter rated at a maximum output power not listed above, the recommended separation distance in meters (m) can be
estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the
transmitter in watts (w) according to the transmitter manufacturer.
NOTE:
1) At 80 MHz and 800MHz, the separation distance for the higher frequency range applies.
2) These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from
structures, objects and people.
10-28
Veterinary Application
Содержание NewTom 5G Vet
Страница 1: ...0051 97050812 Rev 1 28 10 2016 ATTACHMENT to the User Manual Veterinary Application Attachment...
Страница 38: ...10 APPENDIX A Technical references 10 8 Veterinary Application...
Страница 39: ...10 APPENDIX A Technical references Veterinary Application 10 9...
Страница 40: ...10 APPENDIX A Technical references 10 10 Veterinary Application...
Страница 41: ...10 APPENDIX A Technical references Veterinary Application 10 11...