![CEFLA MOCOM Supreme-17 Скачать руководство пользователя страница 94](http://html1.mh-extra.com/html/cefla/mocom-supreme-17/mocom-supreme-17_manual_2578944094.webp)
94
EN
CODE
ALARM DESCRIPTION
MESSAGE ON THE DISPLAY
RESET MODE
E061
The autoclave cannot connect to Wi-Fi network
WI-FI CONFIG. ERROR
CHECK SETTINGS
1
E900
Vacuum test failed
(during TEST PHASE)
TEST FAILED
SECOND STEP
CONTACT TECH.SERVICE
2
E901
Vacuum test failed
(during STAND-BY PHASE)
TEST FAILED
FIRST STEP
CONTACT TECH.SERVICE
2
E902
Vacuum test failed
(vacuum pulse time-out exceeded)
TEST FAILED
VACUUM NOT ACHIEVED
CONTACT TECH.SERVICE
2
E998
Remote maintenance activity in progress
REMOTE SERVICE
ACTIVE
1
E999
Manually interrupting the cycle
MANUAL INTERRUPTION
2
1 = OK (warning)
2 = OK + door unl RESET (if in cycle)