5.4.1.
INSTALLING THE RECIRCULATION FILTER INSIDE THE TANK
Take the recirculation filter from the package, then proceed as follows:
1
Turn the sterilizer off;
2
Move the filter locking lever (1) backward;
3
Fit the recirculation filter (2) onto the hole located in the upper part of
the discharge tank;
4
Move the locking lever (1) forward until it reaches the upper part of
the filter;
5
The sterilizer is ready to reuse the discharge water.
Both tanks must be emptied from water before leaving the
sterilizer off for a long period of time.
Failure to empty them will cause the degradation of the
water quality and the proliferation of bacteria.
Due to water stagnation in the demineralisation and
recirculation filter ducts and in the
sterilizer’s tanks.
If the recirculation filter is removed and no longer used, fit in its place the
union (3) supplied with the sterilizer, following the same procedure used
to fit the recirculation filter inside the tank.