CODE
POSSIBLE CAUSE
SUGGESTED SOLUTION
Problem in the hydraulic circuit.
Contact Technical Service
(see Appendix).
E902
Excessive humidity in the sterilization chamber.
RESET according to the instructions.
Thoroughly dry the inside of the chamber and restart the program.
Air seepage through the gasket.
RESET according to the instructions.
Thoroughly clean the gasket with a clean cotton cloth moistened with water.
Restart the program.
Vacuum pump failure.
Contact Technical Service
(see Appendix).
Problem in the hydraulic circuit.
E998
Service maintenance in progress.
Service maintenance in progress. If you were not informed, contact
IMMEDIATELY the manager of the network to which the sterilizer is
connected.
Contact Technical Service (see Appendix).
E999
Manual interruption of the sterilization or test cycle.
RESET according to the instructions.