background image

 

 

 

 

 

 

EN 

    77   

 

19.6. SUMMARY TABLE OF 28 120 V CYCLES  

CYCLE 

DESCRIPTION 

NOMINAL VALUES 

BASIC CYCLE 

PARAMETERS 

STERILIZABLE MATERIALS 

NOTES 

Temp

e

ra

tu

re

 (

°C)

 

P

re

s

s

u

re

 (

b

a

r)

 

R

e

te

n

tion

 t

im

e

 (

mi

n

C

y

c

le

 t

y

p

e

 (

E

N

 1

3

0

6

0

:2

0

1

4

P

re

-v

a

c

u

u

(F=

fr

a

c

tion

a

te

d

S

=

s

in

g

le)

 

S

ta

n

d

a

rd

 d

ry

ing

 (

mi

n

Total 

c

y

c

le

 t

ime 

(Max

 f

ill

ing

Max

 H

2

c

o

n

s

u

mp

tion

 (

ml

/c

y

c

le)

 

A

v

e

ra

g

e

 

e

n

e

rg

y

 

c

o

n

s

u

m

p

tion

 

(k

W

h

/c

y

c

le)

 

TY

P

E

 

MA

X

 TO

TA

L

 M

A

S

S

 (

k

g

MA

X

 M

A

S

S

 P

E

R

 TR

A

Y

 (

k

g

MA

X

 M

A

S

S

 P

E

R

 I

TE

(k

g

 

135°C 

HOLLOW 

WRAPPED 

135 

2.2 

4(*) 

17 

65 

900 

0.8 

Unwrapped porous 

materials 

1.50 

0.50 

0.50 

For wrapped 

materials and 

instruments (single 

and double pack), it 

is advisable to use 

the 3-tray 

configuration 

Porous materials in single 

pack 

1.25 

0.35 

0.35 

Porous materials in 

double pack 

0.90 

0.30 

0.30 

Solid and hollow materials 

in single pack 

5.00 

1.50 

0.75 

Unwrapped solid and 

hollow materials 

9.00 

1.40 

0.25 

Solid and hollow 

instruments in double 

pack 

2.50 

0.70 

0.25 

135°C SOLID 

UNWRAPPED 

135 

2.2 

4(*) 

38 

500 

0.6 

Unwrapped solid and 

hollow materials "B" 

9.00 

1.50 

0.50 

 

121°C 

RUBBER & 

PLASTIC 

121 

1.1 

20 

17 

80 

950 

0.8 

Unwrapped porous 

materials 

1.50 

0.50 

0.50 

For wrapped 

materials and 

instruments (single 

and double pack), it 

is advisable to use 

the 3-tray 

configuration 

Porous materials in single 

pack 

1.25 

0.35 

0.35 

Porous materials in 

double pack 

0.90 

0.30 

0.30 

Solid and hollow materials 

in single pack 

5.00 

1.50 

0.75 

Unwrapped solid and 

hollow materials 

9.00 

1.40 

0.25 

Solid and hollow 

instruments in double 

pack 

2.50 

0.70 

0.25 

135° HOLLOW 

UNWRAPPED 

135 

2.2 

4(*) 

50 

950 

0.7 

Unwrapped hollow 

instruments 

9.00 

1.50 

0.50 

 

Unwrapped solid 

instruments 

9.00 

1.50 

0.50 

135°C SOLID 

WRAPPED 

135 

2.2 

4(*) 

17 

52 

500 

0.6 

Solid and hollow 

instruments "B" in single 

pack 

5.00 

1.00 

0.25 

It is advisable to 

use the 3-tray 

configuration 

Unwrapped solid and 

hollow materials "B" 

9.00 

1.20 

0.50 

 

XXX°C USER 

(see note) 

135-

121 

2.2-

1.1 

4÷30 

20÷30 

n.a. 

5÷30  n.a.  n.a. 

n.a. 

Unwrapped solid 

instruments (other load 

types are possible 

depending on the user 

settings) 

n.a. 

n.a. 

n.a. 

Variable 

parameters 

depending on the 

settings made 

https://stomshop.pro/

Содержание MOCOM Supreme-17

Страница 1: ...97055023 Rev 00 2019 03 Supreme 17 Supreme 22 Supreme 28 EN https stomshop pro...

Страница 2: ...1 4 5 ELECTRICAL CONNECTIONS 22 4 6 DIRECT CONNECTION TO A CENTRALIZED DRAINING POINT 22 5 FIRST START UP 23 5 1 STARTING 23 5 2 MAIN MENU 24 5 3 UPPER DOOR OPENING 24 5 4 DISCHARGE WATER RECOVERY AND...

Страница 3: ...0 V CYCLES 73 19 5 SUMMARY TABLE OF 28 220 V 240 V CYCLES 75 19 6 SUMMARY TABLE OF 28 120 V CYCLES 77 19 7 STERILIZATION PROGRAM DIAGRAM 80 19 8 DIAGRAMS OF THE TEST PROGRAMMES 82 20 APPENDIX MAINTENA...

Страница 4: ...5 1 ERRORS CATEGORY E 99 24 5 2 ALARMS CATEGORY A 102 24 5 3 HAZARDS CATEGORY H 105 24 5 4 SYSTEM ERRORS CATEGORY S 106 25 USER PIN RESET 108 26 APPENDIX ACCESSORIES 108 27 LOCAL PRINTER CONNECTION 11...

Страница 5: ...to high temperature Disposal symbol in accordance with Directive 2012 19 EU Device in compliance with essential requirements of Directive 93 42 EU and subsequent modifications Notified body IMQ spa Re...

Страница 6: ...responsible for any breakage malfunction property damage or injury to people in the event that the product is not installed or used correctly or proper maintenance is not carried out Please observe th...

Страница 7: ...tely separate the materials to be sterilized from those that must not be subjected to such process or are not able to withstand it Contamination due to unsterilised material caused by wrong final asse...

Страница 8: ...atches the specifications of the order see the delivery note There is no visible damage to the product Dimensions and weight A Height 600 mm B Width 600 mm C Depth 700 mm Total weight 68 kg In case of...

Страница 9: ...s EC Declaration of Conformity 11 Rear spacers 4 Lubricant for door locking mechanism 12 USB key containing User manual and DataSter Software 5 Tray extractor 13 Filling tank demineralisation filter...

Страница 10: ...similar means We recommend to transport and store the device at a temperature not below 5 C Extended exposure to low temperatures may damage the product Store the original package and use it for any...

Страница 11: ...and a perfect result The device offers users 6 sterilization programs one of which completely programmable all equipped with customisable optimised drying for the fast effective sterilization of the v...

Страница 12: ...100 m max 3000 m a s l Pressure min 980 hPA max 1045 hPA a s l External dimensions HxWxD rear connections excluded 500 x 480 x 600 mm Net weight unladen unladen with tray holder support and trays unla...

Страница 13: ...0V 60Hz Main power cord Plug NEMA 6 15P 250V 15A SJT 14 AWG 3C 300V 60 C Connector C19 acc to IEC 60320 Ethernet connection RJ45 max cable length 29 m Wi Fi 802 11 b g n 2 4 Ghz WEP WPA WPA2 PSK encry...

Страница 14: ...ction cut off of the electricity to the chamber heating element Door position safety microswitch Confirmation of the correct closing position of the door of the container under pressure Action signall...

Страница 15: ...D 0 05 mg l 0 05 mg l HEAVY METAL RESIDUES iron cadmium and lead excluded 0 1 mg l 0 1 mg l CHLORIDES 2 mg l 0 1 mg l PHOSPHATES 0 5 mg l 0 1 mg l CONDUCTIVITY AT 20 C 15 S cm 3 S cm pH VALUE 5 7 5 7...

Страница 16: ...st filter 5 USB port 6 NFC reader 1 Sterilization chamber 2 Bacteriological filter 3 Door locking system 4 Steam diffuser 5 Water drainage filter 6 Water top up quick connector 7 Water drainage quick...

Страница 17: ...PURE 100 500 and automatic filling accessory kit AUX EV Kit 5 Data plate SERIAL NUMBER LABEL See image 6 Ethernet cable connection max length 29 m 7 Serial cable connection 8 Automatic filling electri...

Страница 18: ...the same as shown on the sterilizer display A Door opening B Settings C Sterilizer information D Performed cycle list E Sterilization cycles F Test cycles G Custom cycle customisable Other particular...

Страница 19: ...ization in the example 134 C 4 Stabilizing the pressure and temperature 5 Sterilizing for the required time in the example 4 minutes 6 Depressurizing the sterilization chamber 7 Vacuum drying phase 8...

Страница 20: ...ions contained hereafter in this chapter Technical service department see appendix is available for any doubt or further information The sterilizer is placed on the marked only after having passed all...

Страница 21: ...echnical characteristics for complete technical data 4 3 GENERAL PRECAUTIONS FOR INSTALLATION To ensure a correct operation of the device and or avoid risk situations respect the following warnings In...

Страница 22: ...ter Remove the cap holding clip and the cap on the rear of the autoclave Fit the plastic tube on the elbow union supplied Fit the union and then refit the clip Fasten the clamp supplied to the drain s...

Страница 23: ...he main switch on the right hand side of the machine Do not turn on the sterilizer if USB key is inserted When the device is first turned on the display shows the selection of LANGUAGE DATE and TIME s...

Страница 24: ...24 EN 5 2 MAIN MENU At the end of starting procedure the main menu is displayed on the side 5 3 UPPER DOOR OPENING Remove the upper panel by lifting it and sliding it forward https stomshop pro...

Страница 25: ...reused to generate steam This new feature allows saving up to 90 water as compared with a standard class B autoclave All this with the objective of offering a product suitable for the daily needs in...

Страница 26: ...The sterilizer is ready to reuse the discharge water Both tanks must be emptied from water before leaving the sterilizer off for a long period of time Failure to empty them will cause the degradation...

Страница 27: ...l a clack is heard indicating that the filter is correctly fitted inside the tank 4 Fill the tank with water from the mains 5 The sterilizer is ready for use 5 4 3 SAFETY PRECAUTIONS Below are the ind...

Страница 28: ...O conductivity at filter outlet B Litres of treated water 5 5 MAINTENANCE OF THE INTEGRATED DEMINERALISATION FILTER The integrated demineralisation filter is a consumable part a message is displayed w...

Страница 29: ...l CHLORIDES 2 mg l 0 1 mg l PHOSPHATES 0 5 mg l 0 1 mg l CONDUCTIVITY AT 20 C 15 S cm 3 S cm pH VALUE 5 7 5 7 ASPECT colourless transparent without sediment colourless transparent without sediment HAR...

Страница 30: ...le in user friendly menus Use the configuration program whenever necessary A correct customisation of the device provides the best performance and the most satisfactory use The technical service depar...

Страница 31: ...EN 31 6 1 2 DATE AND TIME Select DATE AND TIME option Select the field to edit Set DATE and TIME and confirm by pressing the CONFIRM icon https stomshop pro...

Страница 32: ...f temperature and pressure values Confirm by pressing the CONFIRM icon 6 1 4 SCREEN AND AUDIO Select SCREEN AND AUDIO to set the parameters of screensaver audio volume and brightness of screen and LED...

Страница 33: ...screensavers provided or to load an image from an USB memory support to be connected to the front USB port Use the AUDIO control to set the volume of notifications Use the BRIGHTNESS control to set th...

Страница 34: ...brightness value is set to 00 both bars remain off 6 1 5 PREHEATING Select PREHEATING to activate the PREHEATING of the sterilization chamber Select ACTIVE to activate the PREHEATING OFF to disable i...

Страница 35: ...e actually connected by pressing the corresponding key If connecting the filling system when the sterilizer is off access the menu via the configuration program and manually select the correct option...

Страница 36: ...end of the cycle LABELS label printing activated including barecodes cycle data and device data REPORT printing of a compact version of the cycle summary report activated EXTENDED REPORT printing of...

Страница 37: ...ls option is selected the dedicated menu is displayed when the door opens at the end of the cycle to allow setting Number of labels Package validity interval in days The date that will be printed is i...

Страница 38: ...e screen From CYCLE LIST it is possible to download the reports stored in the device on a USB memory support inserted in the front USB port The download can be customised by using the selection boxes...

Страница 39: ...control activates the download of reports based on the filters enabled Reports are downloaded in PDF format The control activates the printing of cycle reports according to the mode selected https sto...

Страница 40: ...WIFI Select Wi Fi to connect the sterilizer to a local Wi Fi network DHCP can be set as automatic or manual In automatic mode the network configuration parameters are assigned automatically in manual...

Страница 41: ...it is possible to set DHCP as automatic or manual If automatic mode is selected the network configuration parameters are assigned automatically whereas in manual mode they must be manually set 8 4 NF...

Страница 42: ...SET After you have entered the PIN you can access the reserved administrator menu The ADMIN user can decide whether the sterilizer will request the PIN to the generic user at the cycle start PIN Start...

Страница 43: ...possible to enter up to 30 users in the list 9 2 CHANGE This function reserved to ADMIN allows associating one or more sterilization cycles with each user choosing them among the six available ones Ea...

Страница 44: ...44 EN 9 3 CHANGE PIN This function allows each single user to change the identification PIN originally entered https stomshop pro...

Страница 45: ...llows activating the user identification request Only at CYCLE START Only at CYCLE END At CYCLE START and END If the PIN is requested at CYCLE START and END the system accepts the identification of tw...

Страница 46: ...on is not reversible The name of the deleted user will still be shown in any cycle reports associated with the removed account Select ADMIN user to view the summary data on the user The ADMIN user can...

Страница 47: ...sonal PIN You will access the screen containing data of the selected user Once entered a NON ADMIN user can see only a summary of his her data or change his her PIN see entering PIN the following is r...

Страница 48: ...ons and video tutorials for the user Press on the video of your interest to show the selected tutorial on the sterilizer display 11 SERVICE This menu is intended for the technical service department I...

Страница 49: ...must only be applied when it is strictly necessary Solutions containing phenols or quaternary ammonia compounds can cause corrosion on instruments and on the metal parts of the ultrasound device 4 Af...

Страница 50: ...bes etc resting on their side or upended thus avoiding pooling water Do not load trays beyond the limit indicated see Appendix Do not stack trays one on top of the other and do not put them in direct...

Страница 51: ...if necessary If possible using a suitable support position bags at right angles with tray Never stack bags one on top of the other Always pack instruments if they have to be stored for a long time Ref...

Страница 52: ...6 compartments based on the sterilizer model Or if tray holder support is extracted and turned by 90 it can be used to house special boxes 3 4 compartments based on sterilizer model In any case it is...

Страница 53: ...he nature of the problem With this type of control when you select a suitable sterilization program you are guaranteed an effective sterilization under any conditions After inserting the load in the s...

Страница 54: ...allows accessing the start delay setting expressed in hours The delayed start date and time are indicated at the centre The cycle start screen and the set delay value will be displayed when confirmin...

Страница 55: ...4 C UNIVERSAL program characterised by fractionated pre vacuum the cycle sequence is as follows WARMING UP FIRST VACUUM PHASE FIRST PRESSURE RISE SECOND VACUUM PHASE SECOND PRESSURE RISE THIRD VACUUM...

Страница 56: ...o open the sterilizer door press the icon indicated in the figure User identification is required in this phase if enabled If enabled the label printing management screen will be displayed once the do...

Страница 57: ...of pre vacuum the process temperature the exposure time and extra drying if necessary Once selections have been made save the settings with CONFIRM key Press INFO to display the information on the cyc...

Страница 58: ...before any transport and subsequent storage If necessary for transport transfer the material using dry clean and disinfected containers The containers must be closed or if open covered with clean clo...

Страница 59: ...the sterilization chamber tubes and the various interception devices is To select VACUUM TEST cycle press the corresponding icon The cycle must be run with the sterilization chamber empty and only the...

Страница 60: ...r Close the door and start the cycle The test cycle takes place with a succession of phases similar to those described for a normal sterilization cycle At the end of the cycle remove the test device f...

Страница 61: ...ick Test Vacuum Helix Bowie Dick Test based on the available options Once the fields have been set activate the reminder with CONFIRM Reminders activate at 8 a m of the selected day or upon switching...

Страница 62: ...62 EN 15 5 H2O TEST Select this option to test the water quality The water conductivity is automatically measured at each sterilization or test cycle start https stomshop pro...

Страница 63: ...ince the cycle could not finish correctly In certain phases of the cycle the manual interruption activates an automatic cleaning procedure of the internal hydraulic circuit that can last some minutes...

Страница 64: ...free end of the drain tube provided in the basin 2 Insert the other end of the tube in the female union beneath the chamber inlet connector on the left pushing down until you hear a click 3 Completely...

Страница 65: ...ta about the cycles carried out stored in the inner memory of the sterilizer on a USB key To download the files of sterilization cycles tests select the following button USB key has to be formatted ac...

Страница 66: ...gauze pads etc Other generic material suitable for autoclave treatment Depending on the material conformation solid hollow or porous on any package containing it paper plastic bag paper for sterilizat...

Страница 67: ...60 0 20 0 20 Solid and hollow materials in single pack 3 00 1 00 0 50 Unwrapped solid and hollow materials 6 00 1 20 0 25 Solid and hollow instruments in double pack 1 50 0 50 0 25 134 C PRION 134 2...

Страница 68: ...0 1 20 0 50 Unwrapped solid instruments 6 00 1 20 0 50 134 C SOLID WRAPPED 134 2 1 4 S S 13 31 350 0 55 Solid and hollow instruments B in single pack 3 00 1 00 0 25 It is advisable to use the 3 tray c...

Страница 69: ...00 1 20 0 25 Solid and hollow instruments in double pack 1 50 0 50 0 25 135 C SOLID UNWRAPPED 135 2 2 4 S S 4 26 350 0 55 Unwrapped solid and hollow materials B 6 00 1 20 0 50 121 C RUBBER PLASTIC 121...

Страница 70: ...nergy consumption kWh cycle TYPE MAX TOTAL MASS kg MAX MASS PER TRAY kg MAX MASS PER ITEM kg XXX C USER see note 135 121 2 2 1 1 4 30 20 30 n a F 5 30 n a n a n a Unwrapped solid instruments other loa...

Страница 71: ...0 75 0 25 0 20 Solid and hollow materials in single pack 4 00 1 25 0 50 Unwrapped solid and hollow materials 7 50 1 20 0 25 Solid and hollow instruments in double pack 2 00 0 60 0 25 134 C PRION 134 2...

Страница 72: ...0 1 50 0 50 Unwrapped solid instruments 7 50 1 50 0 50 134 C SOLID WRAPPED 134 2 1 4 S S 15 37 400 0 6 Solid and hollow instruments B in single pack 4 00 1 00 0 25 It is advisable to use the 3 tray co...

Страница 73: ...50 1 20 0 25 Solid and hollow instruments in double pack 2 00 0 60 0 25 135 C SOLID UNWRAPPED 135 2 2 4 S S 5 32 400 0 6 Unwrapped solid and hollow materials B 7 50 1 50 0 50 121 C RUBBER PLASTIC 121...

Страница 74: ...nergy consumption kWh cycle TYPE MAX TOTAL MASS kg MAX MASS PER TRAY kg MAX MASS PER ITEM kg XXX C USER see note 135 121 2 2 1 1 4 30 20 30 n a F 5 30 n a n a n a Unwrapped solid instruments other loa...

Страница 75: ...0 90 0 30 0 30 Solid and hollow materials in single pack 5 00 1 50 0 75 Unwrapped solid and hollow materials 9 00 1 40 0 25 Solid and hollow instruments in double pack 2 50 0 70 0 25 134 C PRION 134...

Страница 76: ...0 1 50 0 50 Unwrapped solid instruments 9 00 1 50 0 50 134 C SOLID WRAPPED 134 2 1 4 S S 17 43 500 0 7 Solid and hollow instruments B in single pack 5 00 1 00 0 25 It is advisable to use the 3 tray co...

Страница 77: ...38 500 0 6 Unwrapped solid and hollow materials B 9 00 1 50 0 50 121 C RUBBER PLASTIC 121 1 1 20 B F 17 80 950 0 8 Unwrapped porous materials 1 50 0 50 0 50 For wrapped materials and instruments singl...

Страница 78: ...ad VACUUM TEST 0 8 18 Empty chamber VACUUM HELIX BD TEST executable in sequence 50 To set a sterilization time of 5 5 minutes contact the Technical Service Single Pre Vacuum 1 pre vacuum 0 8 bar see f...

Страница 79: ...81 0 79 t4 2PP 0 97 1 03 t5 3PV 0 81 0 79 t6 SS 4 5 5 134 138 2 04 2 40 t7 SE 4 5 5 134 138 2 04 2 40 t8 DS 0 81 0 79 t9 DE 2 CE 0 02 0 02 121 C cycles EN 13060 2014 Time minutes Min temperature Max t...

Страница 80: ...NIVERSAL 134 C 4 00 PROGRAM 134 C PRION 134 C 18 00 A PRESSURE BAR B PROCESS C TIME MIN D FRACTIONATED VACUUM E VACUUM DRYING PROGRAM 121 C UNIVERSAL 121 C 20 00 A PRESSURE BAR B PROCESS C TIME MIN F...

Страница 81: ...UNWRAPPED 134 C 4 00 A PRESSURE BAR B PROCESS C TIME MIN F FRACTIONATED PRE VACUUM H SHORT DRYING PROGRAM 134 C SOLID WRAPPED 134 C 4 00 A PRESSURE BAR B PROCESS C TIME MIN I SINGLE PRE VACUUM G LONG...

Страница 82: ...A PRESSURE BAR B PROCESS C TIME MIN F FRACTIONATED PRE VACUUM H SHORT DRYING PROGRAM VACUUM TEST 0 80 bar A PRESSURE BAR C TIME MIN L INTERMEDIATE CONDITION TO CONTINUE THE TEST M FINAL CONDITION TO P...

Страница 83: ...x The ordinary maintenance described below consists in easy manual operations and preventive interventions involving simple instruments In the event of replacement of components or parts of the device...

Страница 84: ...ollowing general warnings Do not wash the sterilizer with direct jets of water neither under pressure nor sprinkled Seepage into electrical and electronic components could damage the functioning of th...

Страница 85: ...d with 50 water Apply product with a soaked cloth then dry As an alternative we recommend using products containing the following at no more than the given concentration Ethanol Concentration maximum...

Страница 86: ...n filter maintenance is due or every time you notice visible clogging of the filter indicated by the filter markedly turning grey unscrew the bacteriological filter B from its support and replace it w...

Страница 87: ...lifting the rear part then the front part Replace the cartridge following the initial installation procedure see INTEGRATED DEMINERALISATION SYSTEM 20 3 10 RECIRCULATION SYSTEM CARTRIDGE REPLACEMENT R...

Страница 88: ...y information relative to the periodic validation of steam sterilizers 20 5 DEVICE USEFUL LIFE Water steam sterilizer service life is of 10 years average use 5 cycles day for 220 days year For normal...

Страница 89: ...e sterilizer to reach the proper operating conditions for starting the program Under standard conditions the Average Preheating Time is about 10 15 Minutes The safety valve has triggered Locking ring...

Страница 90: ...s on the instruments Carefully clean the material before subjecting it to the sterilization cycle See Chapter Preparing the material Contact between instruments made of different metals Separate instr...

Страница 91: ...a specialised technician Code format Axxx xxx identification number 000 999 H HAZARD Second level fault The problem can normally be solved by the Technical Service Centre Code format Hxxx xxx identif...

Страница 92: ...e OK button 2 Following the instructions displayed and holding down the RESET button for about 3 seconds Press the RESET button for approx 3 seconds to go back to the main menu After the RESET and any...

Страница 93: ...QUALITY H2O BAD CHANGE FILTERS CHANGE WATER 2 E012 The cycle threshold for the periodic replacement of integrated water filters has been reached CHANGING FILTERS 1 E013 Water conductivity threshold 2...

Страница 94: ...TACT TECH SERVICE 2 E901 Vacuum test failed during STAND BY PHASE TEST FAILED FIRST STEP CONTACT TECH SERVICE 2 E902 Vacuum test failed vacuum pulse time out exceeded TEST FAILED VACUUM NOT ACHIEVED C...

Страница 95: ...ed sterilization chamber CHAMBER PROBE PT1 SHORT CIRCUIT CONTACT TECH SERVICE 1 A112 Temperature sensor PT2 short circuited steam generator GENERATOR PROBE PT2 SHORT CIRCUIT CONTACT TECH SERVICE 1 A11...

Страница 96: ...2 A254 2nd rise back up to atmospheric pressure not reached within the time out ATM2 UPSTROKE TIMEOUT CONTACT TECH SERVICE 2 A255 2nd pressure pulse not reached within the time out 2PP UPSTROKE TIMEOU...

Страница 97: ...OW MINIMUM LIMIT CONTACT TECH SERVICE 2 H404 Floating temperature over the limit STERILIZATION phase ERRATIC TEMPERATURE CONTACT TECH SERVICE 2 H405 Pressure over the MAX limit STERILIZATION phase PRE...

Страница 98: ...AD NEW CYCLES 2 S021 Cycle storage limit exceeded CYCLE MEMORY FULL START OVERWRITING 2 S030 Check using a watchdog that one of main tasks is not in crash condition SYSTEM ERROR CONTACT TECH SERVICE 2...

Страница 99: ...t Technical Service see Appendix Disturbance on the electrical mains RESET according to the instructions If the problem occurs again have the electrical mains checked by a technician If the electrical...

Страница 100: ...ice see Appendix E031 Water level in the drain tank over the MAX level RESET according to the instructions and empty the tank Completely drain the drain tank MAX water level sensor failure Contact Tec...

Страница 101: ...tructions Thoroughly clean the gasket with a clean cotton cloth moistened with water Restart the program Vacuum pump failure Contact Technical Service see Appendix Problem in the hydraulic circuit E99...

Страница 102: ...5 failed conductivity measurement compensation A111 Incorrect temperature sensor connection sterilization chamber Temperature sensor short circuit sterilization chamber A112 Incorrect temperature sens...

Страница 103: ...ed with water or in general with liquids into the chamber Air seepage through the gasket RESET according to the instructions Thoroughly clean the gasket with a clean cotton cloth moistened with water...

Страница 104: ...ilter obstructed Clean the drain filter see Maintenance Appendix Problem in the hydraulic circuit Contact Technical Service see Appendix A262 Bacteriological filter obstructed Clean the drain filter s...

Страница 105: ...aulic circuit H402 Steam generator malfunction Problem in the hydraulic circuit H403 Steam generator malfunction Problem in the hydraulic circuit H404 Problem in the hydraulic circuit Steam generator...

Страница 106: ...ake sure that printer lid is correctly closed Check printer cable correct connection If the problem persists contact Technical Service see Appendix S012 Printer not ready for use Make sure that paper...

Страница 107: ...trol board or software Contact Technical Service see Appendix S099 Malfunction of control board or control software Try restarting the program a second time Try replacing the USB key If the problem pe...

Страница 108: ...eet the manufacturer s specifications DEMINERALIZER PURE 100 DEMINERALIZER PURE 500 TWIN PURE 500 AUTOMATIC FILLING FRONT FILLING H2O AUXILIARY SOLENOID VALVE Additional SV kit including 1 2 way water...

Страница 109: ...EN 109 EXTERNAL PRINTER BARCODE READER NETWORK PRINTER DATA STER SOFTWARE MY TRACE SOFTWARE https stomshop pro...

Страница 110: ...tivated only if the NETWORK PRINTER and the sterilizers are connected to the same LAN Local Area Network through a Wi Fi connection or an Ethernet cable For a correct installation and configuration of...

Страница 111: ...meet the manufacturer s specifications Description Code bacteriological filter 97290160 door gasket 17 22 l 97400145 door gasket 28 l only 97467176 demineralised water tank chamber filter 97290210 dis...

Страница 112: ...water steam sterilization procedures In this regard please refer to the following address Cefla S c Plant Via Bicocca 14 C 40026 Imola BO IT Tel 39 0542 653441 Fax 39 0542 653555 Headquarters Via Sel...

Страница 113: ...www cefla com https stomshop pro...

Отзывы: