page 7
page 6
TORQUE MEASUREMENT
TORQUE MEASUREMENT
6.
Beeper (bp)
After Auto Zero Reset is entered, "bp" is displayed, then "On".
Press UP or DOWN/SHIFT to select On, OFF, or FF, then press
PRG/SET to enter.
"On"– beeper sounds for Good, NG and capacity overload
"OFF"– beeper sounds for capacity overload only
"FF"–
beeper sounds for NG and capacity overload.
After "bp" is entered, "-S-" is displayed to confirm programming com-
pletion and zero is shown.
After High and Low setpoints are set and beeper is set to On,
In PP mode
Under LO setpoint– No LED light or beep
Over LO setpoint–
Green LED lights and beep sounds
Over HI setpoint–
NG flashes and beep sounds
In PD mode– no indicator or beep for values before Peak
down is sensed.
Under LO setpoint– NG flashes and slow beep sounds
Between LOand HI– GOOD lights and constant tone sounds
Over HI setpoint–
NG flashes and fast beep sounds
In Real Time mode
Under LO setpoint– no LED light or beep
Between LO and HI–GOOD lights and beep sounds
Over HI setpoint–
NG lights solid and beep sounds
DATA MEMORY
Storing and Recalling Data from Memory
Memory functions work in PP and PD modes. Store up to 800 val-
ues in memory.
1. Measure in PP or PD mode, when the display resets by either
Auto Zero or manually pressing CLEAR, the peak value is stored.
2. To recall a value, press MEMORY and the last stored memory is
displayed. Press UP or DOWN/SHIFT to select a memory loca-
tion (i.e. .0.0.1 for first location) and torque value. The display
cycles between memory location and value.
Clearing Data from Memory
Single clear:
Press MEMORY, move to a memory location and
when the desired memory value is displayed, press CLEAR, "CLR" is
displayed. Press CLEAR again to clear the data and "– – – –" is dis-
played to confirm deletion.
All clear:
Press CLEAR and hold until the display shows "ALL" then
press CLEAR again and "– – – –" is displayed to confirm deletion.
Downloading Memory Data
1. Press ON/OFF to turn on.
2. Press MEMORY and after memory data is displayed press MEMORY
again and "FA" is displayed. Use UP or DOWN/SHIFT to select the
first memory location then press MEMORY and "LA" is displayed.
Use UP or DOWN/SHIFT to select the last memory location.
Press MEMORY again and data is sent. While data is output -P- is
displayed.
Statistics
Press STATISTICS and the number of data, Max, Min, and Avg are dis-
played in sequence.
Auto Power Off
To maximize the life of the battery, power automatically shuts off
after 10 minutes of non-use.
Low Battery Indicator
Battery indicator status shows full, half or needs recharging.
If the battery is empty, power is turned off immediately.
IMPORTANT!
Use the provided CEDAR AC adapter/charger
exclusively and plug into the correct AC output. It takes 3
hours to fully recharge for 12 hours of continuous use. When fully
charged, the charge light goes off.
System Reset
When battery power is completely depleted, the tester may not
work even though it has been recharged. In this case, press the
System Reset Button.